9 Troubleshooting
Fault
Cause
Solution
irregular flow of
blasting material
too little blasting material in
container
re-fill the container
nozzle on stylus too small
use larger nozzle
blasting material is moist
use blasting material of correct dryness
no flow of blasting
material
container empty
re-fill the container
main nozzle and/or steel
hose blocked
•
use recommended, dry blasting
material
•
unscrew nozzle on the stylus
trigger blasting process
blasting material should flow
•
if necessary, build up
counterpressure by pressing the
hose
•
clean the nozzle (make the stuck
grains come loose)
•
if the obstruction persists, use
compressed air or a needle to
clean the nozzle
wrong blasting container
please switch to the correct blasting
container
chamber nozzle blocked
Contact our Service:
•
item no. 190494
(Key for chamber nozzle)
is necessary for dismounting
•
clean the nozzle (make the stuck
grains come loose)
•
if the obstruction persists, use
compressed air or a needle to
clean the nozzle
too little pressurised air < 3
bar
increase pressure
decreasing and/or
to low blasting
pressure
protective filter (item no.
610039 Small Filter Plate
D14 X 1.5) blocked and/or
buckled closed
change filter plate
item no. 610039
Small Filter Plate D14 X 1.5
If the above recommendations do not solve the problem, contact your dental depot
or our service department.
Remarks:
Obstructions normally occur from soiled or lumply blasting material. If the sand
draws moisture, check the condensator. Please also see page 17, 10.2.8 Water Separator.