When using 25 kg bulks:
Note that while using bulk containers, the cans must be positioned at a height
higher than the device. Optional pad stands are available on request.
Remove the closings from the cans filled with two component silicone and screw
connection hoses onto the cans.
Warning: Screw the colour marked connection hose onto the can filled with
coloured silicone compound
–
Component B - screw on. Do not mix up!
If you
mix up the closings even for a short time, the device must be disassembled and
cleaned.
Place the can, e.g. on the optional pad stand, to make sure that it will be placed
higher than the dosage device (can will be upside down) and insert the connection
hoses with quick closings into the entry on the back of the device.
Very important:
Drill a ventilation hole using a drill or similar tool at the upper side
edge of the can to ensure proper ventilation.
Replacing bottles/cans:
Remove the quick coupling/bottle adaptor from the device by pulling up the lock. Screw the
quick coupling/bottle adaptor onto the new bottle/can and connect as explained above.
Take special care to never confuse the bottles or connection hoses A+B – not even
for a short time – since the silicone within the connections and pipes hardens
instantly! The cleaning procedure is very demanding and may only be performed by
authorised repair companies.
Note that while using bulk containers, the cans must be positioned at a height higher than
the device. Optional pad stands are available on request.
Never use with a hardened mixing cannula!