CL-200A Manual, Rev. 120523 
 
 

 

 

CONTROL DESCRIPTION  

Front panel 

A labeled image of the face panel for one channel of the 

CL-200A

 is shown below. Operating 

components are identified by letters. Refer to this diagram as an aid in identification of the described 
components. 

Power Switch (Item A) 

Supplies power to the 

CL-200A

.  

Meter Selector Switch and Meter (Item B) 

T

HE SELECTOR SWITCH

 selects the parameter displayed on the associated 

METER

. Included 

options are 

Set Temp

Control Temp (T1)

Monitor Temp (T2)

Output Volts, 

and 

Output Amps 

and are described below.  

Set Temp

 reports the adjustment from either the 

SET TEMPERATURE 

control (Item H) or the 

EXT

 

TEMP SET

 BNC (Item Q). Displayed units are °C.  

Control Temp

 reports the actual temperature of the feedback thermistor situated within the 

heater/cooler device being used. Displayed data is informative only and is in units of °C. 

Monitor Temp

 reports the temperature of the sensor thermistor which connects via the 

MONITOR TEMP IN

 BNC (Item F). This provides a convenient means to determine the 

temperature of any point of interest within your configuration. Displayed units are °C. 

Output Volts

 reports the adjustment from either the 

MANUAL VOLTAGE 

control (Item K) or the 

EXT

 

VOLT SET

 BNC (Item T). Displayed units are Volts.   

 

                         

                      

Warner

 Instruments 

Harvard Apparatus Company     

Summary of Contents for CL-200A

Page 1: ...e en nt ts s D Du ua al l C Ch ha an nn ne el l B Bi ip po ol la ar r T Te em mp pe er ra at tu ur re e C Co on nt tr ro ol ll le er r M Mo od de el l C CL L 2 20 00 0A A Warner Instruments 1125 Dixwell Avenue Hamden CT 06514 800 599 4203 203 776 0664 203 776 1278 fax ...

Page 2: ... Voltage Control Item K 7 Peltier Limit LED Item L 8 Freeze Alert LED Item M 8 Heat Cool LED s Item N 8 Current Limit LED Item O 8 Feedback Thermistor Selector Switch Item R 8 Thermistor Open LED Item S 8 Rear panel 9 External Volts Set Item T 9 External Temp Set Item U 9 Heater Cooler input Item W 9 Power Entry Module Item Y 9 Ground Terminals Item X 9 SETUP 10 Warner Peltier Driven Systems 10 Op...

Page 3: ...CL 200A Manual Rev 120523 APPENDIX 13 Specifications 13 Accessories and replacement parts 13 Warranty and service 14 Warranty 14 Service notes 14 Certifications 15 ...

Page 4: ...ple to use with a single control for temperature adjustment Total automatic control is provided in automatic mode while manual control is available in manual mode A loop speed selector is used to optimize the response of the system to accommodate thermal delay characteristics intrinsic to the setup Features of the CL 200A include Quiet operation suitable for use in sensitive electrophysiology appl...

Page 5: ... goal is to provide clarity rather than complexity we welcome any feedback you may wish to provide Warner Instrument product model numbers are presented using bold type References to controls are specified using SMALL CAPS References to control settings are specified using italic type Special comments and warnings are presented in highlighted text Any other formatting should be apparent from conte...

Page 6: ...w Set Temp reports the adjustment from either the SET TEMPERATURE control Item H or the EXT TEMP SET BNC Item Q Displayed units are C Control Temp reports the actual temperature of the feedback thermistor situated within the heater cooler device being used Displayed data is informative only and is in units of C Monitor Temp reports the temperature of the sensor thermistor which connects via the MO...

Page 7: ...CL 200A is used in automatic mode Feedback options include fast medium and slow Optimally set to fast for most applications but can be set to medium or slow for systems with longer thermal delay characteristics For example use the medium speed for use with the SC 20 Bipolar In Line Solution Heater Set Temperature Control Item H This dial adjusts the set point of the automatic temperature control s...

Page 8: ...E LED is lit when the Peltier device within the SC 20 approaches 0 C Activation of this circuit is informative only and will not alter any instrument setting Heat Cool LED s Item N RED and BLUE LED S indicate heating or cooling power applied to the Peltier device Current Limit LED Item O This indicator light informs the user that the channel has reached the maximum current output of 5 Amperes This...

Page 9: ...A Input functions only when the INTERNAL EXTERNAL SELECT CONTROL Item J is set to external and the COOLING MODE SWITCH Item C is set to auto Units are 100 mV C This input can be used for computer control purposes Heater Cooler Peripheral Device Output Item W This is the connection port for Warner Heater Cooler devices such as the SC 20 Dual In line Heater Cooler and QE 1HC Heating Cooling Platform...

Page 10: ...Item U or EXT VOLT SET Item T BNC s on the rear panel Operating With Other Equipment The CL 200A can be used to power other Warner equipment such as the SH 27B and SF 28 inline solution heaters In addition the controller can be used as an independent device to power third party equipment so long as the following considerations are met Warner Resistive Heating Systems Use of the CL 200A to power Wa...

Page 11: ...th feedback thermistors can be used in automatic mode provided they are compatible with the CL 200A requirements LOOP SPEED is normally set to fast to provide the shortest cycle time between the application of power to the heater element and the sensing of temperature at the thermistor For heater cooler thermistor systems with long response times such as with Warner s SC 20 in line solution heater...

Page 12: ...nto manual mode by switching the COOLING MODE SWITCH Item C to manual Switch the METER Item B to Output Volts and adjust to the desired set point using the MANUAL VOLTAGE control Item K Alternatively the CL 200A can be commanded using external inputs Set the CL 200A to external commands by setting the INTERNAL EXTERNAL SELECT control Item J to external In this mode the CL 200A will accept external...

Page 13: ... External Voltage Set BNC Heater Cooler Device 15 pin D connector Temperature Range Peltier 6 C to 65 C Resistive Ambient to 65 C Meter Display Resolution 3 digit LED display of C or V 0 1 C 0 01 Volt 0 01 Ampere Meter Readouts Set temperature C Control Temperature C Monitor Temperature C Output Voltage V Output Current A Enclosure 8 9 X 42 6 X 29 2 cm W x H x D Weight 10 lb 4 54 Kg Operating cond...

Page 14: ...h a 5x20 mm 0 8 A slow blow fuse 0 4 A for facilities using 220 240 V line voltages If the replacement fuse also fails please call Warner Instruments for assistance B Occasionally a knob on the front panel will loosen after long use These are collet style knobs and are tightened with a screw located under the knob cap To gain access to the adjustment screw pry the cap off with a thin bladed screwd...

Page 15: ...0000 4 8 EN610000 4 11 Manufacturer s Name Warner Instruments LLC Manufacturer s Address 1125 Dixwell Avenue Hamden CT 06514 Tel 203 776 0664 Equipment Description Power Controller Equipment Class ITE Class A Model Numbers CL 200A I the undersigned hereby declare that the equipment specified above conforms to the above Directive s and Standard s Place Hamden Connecticut USA Signature Full Name Bur...

Page 16: ...ments LLC Manufacturer s Address 1125 Dixwell Avenue Hamden CT 06514 Tel 203 776 0664 Equipment Description Power Controller Safety requirements for electrical equipment for measurement and laboratory use Equipment Class Class I Model Numbers CL 200A I the undersigned hereby declare that the equipment specified above conforms to the above Directive s and Standard s Place Hamden Connecticut USA Sig...

Page 17: ...CL 200A Manual Rev 120523 Warner Instruments A Harvard Apparatus Company 17 ...

Reviews: