96
MN-111 | Instructions for use | MN-111 W&H Med ENG Rev11 | 2019/01/25 | © 2019 W&H Sterilization Srl
Technical data
Water quality
FEED WATER SPECIFICATIONS (EN 13060)
Notice: Do not use rust inhibitor or any other agents in the clean
water tank.
This sterilizer uses distilled or demineralized water to generate
steam for the sterilization process. The table below lists the
maximum content of minerals and the specifications for the water
used for steam sterilization according to EN13060 ANNEX C.
Contaminants/minerals/qualities
Value/Specification
Evaporate residue
< 10 mg/l
Silicon oxide, SiO2
< 1 mg/l
Iron
< 0.2 mg/l
Cadmium
< 0.005 mg/l
Lead
< 0.05 mg/l
Heavy metals (excl. iron,
cadmium, lead)
< 0.1 mg/l
Chloride
< 2 mg/l
Phosphate
< 0.5 mg/l
Conductivity (at 20°C)
< 15 µs/cm
pH value
5–7
Contaminants/minerals/qualities
Value/Specification
Appearance
colorless, clean, free from sediment
Hardness
< 0.02 mmol/l
Chemical additives
No chemicals or additives must be added to the
water used for the steam sterilization process,
even if they are specifically claimed for use in
steam generators, or for steam production, or
as additives for sterilization, disinfection,
cleaning or corrosion protection.