Step
Description
Key Points
104-001-A.7 Released 06/14/2019
2
CONNECT THE NITROGEN CALIBRATION GAS
2.1
Attach M5 barb fitting
(provided)
to calibration beaker inlet.
The inlet is the side that is closer
to the flange than the other
fitting.
A calibration beaker is included
with the initial analyzer order.
Ensure the provided
O-Ring
is
installed on the fitting to
provide a seal against leaks
between the fitting and the
beaker.
2.2
Install the calibration beaker by
pushing straight upwards over the
sensor.
2.3
Inspect O-Ring
on provided gas
regulator for any debris or
damage.
Take care to ensure there
are no leaks between the
calibration gas tank and the
sensor.