Waltron 9065 Quick Reference Manual Download Page 4

 

Step 

 

Description 

Key Points 

 

 

104-001-A.7 Released 06/14/2019

 

1.5 

 

Disconnect the plumbing

 from the 

flow cell. 

 

1.6 

 

Remove the flow cell. 

 

1.7 

 Air dry 

Allow the sensor to 

air dry

 and to 

allow the sensor and the 
calibration kit to equalize to the 
same temperature.  You may flow 
extra calibration gas or clean dry 
air over the sensor to expedite 
drying. 

 

DO NOT wipe dry the black 
sensor spot.  Touching the 
sensor may damage the 
delicate coating. 

 

1.8 

 

Store the calibration kit in the 
same room as the sensor for at 
least 4 hours

 while the sensor is 

drying to equalize to the same 
room temperature. 

 

The calibration process is 
temperature sensitive.  The 
temperature of the sensor 
must be roughly equal to the 
temperature of the calibration 
gas. 

Summary of Contents for 9065

Page 1: ...104 001 A 7 Released 06 14 2019 9065 Luminescent Dissolved Oxygen Analyzer Calibration Kit Quick Reference Manual...

Page 2: ...fice 908 534 5100 x120 Waltron s technical expertise and extensive experience provides personalized solutions to the water quality industry It is Waltron s commitment to provide the customer with time...

Page 3: ...n K6000 101 A calibration beaker is provided with each 9065 analyzer order Contact Waltron for additional beakers p n K5000 039 1 3 Calibration is Due Determine if the Dissolved Oxygen Analyzer requi...

Page 4: ...flow extra calibration gas or clean dry air over the sensor to expedite drying DO NOT wipe dry the black sensor spot Touching the sensor may damage the delicate coating 1 8 Store the calibration kit...

Page 5: ...itting A calibration beaker is included with the initial analyzer order Ensure the provided O Ring is installed on the fitting to provide a seal against leaks between the fitting and the beaker 2 2 In...

Page 6: ...ey Points 104 001 A 7 Released 06 14 2019 2 4 Remove the protective cap from the provided nitrogen tank 2 5 Install the gas regulator on nitrogen gas tank 2 6 Install one end of the provided tubing on...

Page 7: ...vided tubing on to the barb fitting on the calibration beaker 2 8 Open the regulator all the way to allow flow of Nitrogen gas The provided fixed flow regulator will supply the correct flow rate 2 9 W...

Page 8: ...d sensor in a dual stream analyzer Press ENTER button Go to O2 SENSOR 1 or 2 CALIBRATION Menu If you have a dual stream analyzer you may select Sensor 1 or Sensor 2 3 3 For oxygen free calibration gas...

Page 9: ...4 Using the Arrow buttons select O2 AUTO Press ENTER button In the next screen select O2 LOW Press ENTER button Select 000000 ppb Press ENTER button 3 5 Calibration will begin Wait approximately 2 min...

Page 10: ...Contact Waltron technical support if you are still unsuccessful 3 7 Close the valve on the gas regulator to preserve your calibration gas 3 7 Use the arrow buttons to select STORE CALIBRATION Press EN...

Page 11: ...Replace plastic cap and store in a clean safe location in the provided kit box Remove the calibration beaker and store in a clean safe location for the next calibration Reinstall the flow cell and the...

Reviews: