Step
Description
Key Points
104-001-A.7 Released 06/14/2019
1.5
Disconnect the plumbing
from the
flow cell.
1.6
Remove the flow cell.
1.7
Air dry
Allow the sensor to
air dry
and to
allow the sensor and the
calibration kit to equalize to the
same temperature. You may flow
extra calibration gas or clean dry
air over the sensor to expedite
drying.
DO NOT wipe dry the black
sensor spot. Touching the
sensor may damage the
delicate coating.
1.8
Store the calibration kit in the
same room as the sensor for at
least 4 hours
while the sensor is
drying to equalize to the same
room temperature.
The calibration process is
temperature sensitive. The
temperature of the sensor
must be roughly equal to the
temperature of the calibration
gas.