27
28
100
NA
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation dis-
tance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter,
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
• Guidance and manufacturer’s declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed below, and should only be
used in such environments:
Immunity
test
IEC 60601
test level
Compliance
level
Electromagnetic environment – guidance
Conducted
RF
IEC 61000-
4-6
Radiated
RF
IEC 61000-
4-3
3 Vrms
150 kHz to
80 MHz
3 V/m
80 MHz to
2.5 GHz
Not Ap-
plicable
10 V/m
Portable and mobile RF communications equipment should be
used no closer to any part of the device, including cables, than
the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2
80 MHz to 800 MHz
d = 2 .3
800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey, a should be less than the compli-
ance level in each frequency range.b
Interference may occur in the vicinity of equipment marked
with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
device is used exceeds the applicable RF compliance level above, the device should be observed to verify
normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the device.
b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
TROUBLESHOOTING
If any abnormality arises during use, please check the following points.
Note: If the unit still does not work, contact HoMedics Consumer Relations.
Under no circumstance should you disassemble or attempt to repair
the unit by yourself. Contact information for HoMedics Consumer
Relations Department can be found on the warranty page.
SYMPTOMS
POSSIBLE
CAUSES
CORRECTION
Unit does not turn
on when the
START/
STOP button is
pushed.
Batteries have
run down.
Replace them with
four new “AAA”
alkaline batteries.
Battery polarities
have been positioned
incorrectly.
Re-insert the batteries in
the correct positions.
EE measurement error
symbol shown on
display or the blood
pressure value is
displayed excessively
low (or high).
Cuff has been placed
incorrectly.
Wrap the cuff properly so
that it is positioned
correctly. Measure again.
Did you talk or move
during measurement?
Keep arm steady during
measurement. Measure
again.
Shaking of the arm with
the cuff on.
E1 error symbol shown
on display
Air circuit abnormality.
Cuff tube may not be
plugged into monitor
correctly.
Check cuff connection.
Measure again.
E2 error symbol shown
on display
Inflation pressure
exceeding 300 mmHg.
Switch the unit off, then
measure again.
E3 error symbol shown
on display
Error determining
measurement data.
Measure again.
EP error symbol shown
on display
System error.
Measure again.