CAUTION: There are live circuits inside the controller even when the power switch is in the 

OFF position! Servicing and troubleshooting the controller must be performed by Qualified 

Service Personnel.

 

 

 

Section 1: Electrode Reading Responds Slowly 

 

Step 1: 

 

pH electrodes require periodic cleaning and calibration. 

  The method of cleaning the electrode will depend upon the coating, as well as the 

materials of construction of the electrode. Do not use a solvent that will attack the 

electrode. 

 

  Oily coatings should be removed with a mild detergent or isopropyl alcohol. 

 

  Hard scales such as calcium carbonate can usually be removed with a dilute 

hydrochloric acid solution. 

 

  Soft coatings can be removed using a soft cloth or soft toothbrush. 

 

  A two-point calibration should always be performed after cleaning the electrode. 

 

Step 2: 

 

If cleaning does not speed up the electrode's response 

 

  Replace the electrode. 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for W100

Page 1: ...ads Low or High Versus a Calibrated Handheld Meter Cannot Calibrate Section 4 Electrode Fault Section 5 pH Reading is Unstable Note pH electrodes are like batteries and their output voltage will chang...

Page 2: ...ich means it wires into the controller analog input terminals o The WEL Transmitter Housing only works with the W600 and W900 controllers not with the W100 controller You can determine which configura...

Page 3: ...ill depend upon the coating as well as the materials of construction of the electrode Do not use a solvent that will attack the electrode Oily coatings should be removed with a mild detergent or isopr...

Page 4: ...ith the WEL PH housing wired into the electrode input terminals in the controller and the electrode threaded into the housing the reading should be the pH value of the solution as verified by a second...

Page 5: ...t is programmed as pH and the following configurations are in place With nothing connected to the analog input terminals in the controller the reading should be about 0 mV 7 pH With just the WEL Trans...

Page 6: ...00 for Walchem pH electrodes Temperature element Some electrodes do not have ATC If you are not using ATC in the Inputs menu Temperature Settings scroll to Element and select No Element And whatever t...

Page 7: ...ctrode Input menu look at the pH Raw Value and the Gain Value What is the calibration range that you are allowed to go to o The controller shows us a current pH reading and a raw pH reading o This raw...

Page 8: ...rect using a multimeter please check the mA signal coming from the sensor and into the controller as follows Place the WEL Transmitter Housing plus Electrode in the system tee water or place it in a b...

Page 9: ...condition with stop control of any output using the electrode Possible Cause Corrective Action Electrode wires shorted Disconnect short Faulty electrode Replace electrode Faulty controller Replace or...

Page 10: ...ater does the reading stay stable If the reading does not stay stable this can indicate a ground loop A ground loop could cause readings to be stuck at a high or low value Or there is some electrical...