Maintenance
4.1
Checking and Adjusting
THIS SECTION IS VERY IMPORTANT: Here are a few simple
instructions to check your laser for calibration. The laser is a precision
instrument and it is important that you keep it calibrated and in proper
condition. The accuracy of your work is completely your responsibility
and you should check your instrument before beginning each job, and
especially after the instrument has taken a sharp jolt or been dropped,
or when temperature changes greater than 50 degrees F (28 degrees
C) have occurred.
See graphic: wc_gr004600
The laser has two horizontal axes: X and Y, as indicated on the top of
the laser.
Each end of each axis must be checked for calibration. If needed, the
axis can be calibrated, carefully following the instructions. You can
also take the laser to a Wacker service center for calibration.
wc_gr004600
-X
-X
-Y
-Y
+Y
+Y
+X
+X
Front of laser
Summary of Contents for HAL 300
Page 2: ......
Page 9: ...Wacker HAL 300 Operation wc_tx000803gb fm EN 7 1 3 4 wc_gr004578 8 5 6 11 10 2 7 ...
Page 15: ...Wacker HAL 300 Operation wc_tx000803gb fm EN 13 Notes ...
Page 26: ...Wacker HAL 300 Maintenance wc_tx000806gb fm EN 24 Notes ...
Page 35: ...Wacker HAL 300 Operación wc_tx000803es fm ES 7 1 3 4 wc_gr004578 8 5 6 11 10 2 7 ...
Page 41: ...Wacker HAL 300 Operación wc_tx000803es fm ES 13 Notas ...
Page 52: ...Wacker HAL 300 Mantenimiento wc_tx000806es fm ES 24 Notas ...
Page 62: ...Wacker HAL 300 Fonctionnement wc_tx000803fr fm FR 8 1 3 4 wc_gr004578 8 5 6 11 10 2 7 ...
Page 68: ...Wacker HAL 300 Fonctionnement wc_tx000803fr fm FR 14 Remarques ...
Page 72: ...Wacker HAL 300 Fonctionnement wc_tx000803fr fm FR 18 Remarques ...
Page 80: ...Wacker HAL 300 Maintenance wc_tx000806fr fm FR 26 Remarques ...
Page 83: ......