64
www.nieaf-smitt.com
Operating altitude
2000 meter (7000 feet)
Drop test
2 meters (6.5 feet)
Enclosure
Double molded, waterproof
Category
IV 600 V / III 1000V
Standards
Complies with EN 61010-1 and IEC
61010-1 2
nd
edition (2001), meets UL
61010-1, 2
nd
edition (2004), CAN/CSA
C22.2 No. 61010-1 2
nd
edition (2004), UL
61010B-2-031, 1
st
edition (2003)
Pollution degree
2
Function
Range
Resolution
Accuracy
DC voltage
500 mV
0.01 mV
+ (0.06 % + 4 d)
5 V
0.0001 V
50 V
0.001 V
500 V
0.01 V
1000 V
0.1 V
+ (0.1 % + 5 d)
AC+ DC voltage
50 mV
0.001 mV
+ (1.0 % + 7 d)
50/60 Hz
500 mV
0.01 mV
5 V
0.001 V
50 V
0.01 V
500 V
0.1 V
1000 V
1 V
All AC voltage ranges are specified from 5 % of
range to 100% of range
DC current
500 μA
0.01 μA
+ (1.0 % + 3 d)
5000 μA
0.1 μA
50 mA
0.001 mA
500 mA
0.01 mA
10 A
0.001 A
20 A: 30 sec max with reduced accuracy
AC+DC current
500 μA
0.1 μA
+ (1.5 % + 7 d)
5000 μA
1 μA
50 mA
0.01 mA
500 mA
0.1 mA
50 A
0.01 A
20 A: 30 sec max with reduced accuracy
All AC current ranges are specified from 5 % of
range to 100 % of range
Electrical specificatiions