17
COOLIUS A30
5.2 Used oil weight scale
The display shows:
1. Remove the used oil bottle.
Press
ENTER
key to confirm that the scale is unloaded
and wait!
The display shows:
2. Press
ENTER
key to confirm again that the scale is
unloaded and wait!
The display shows:
3. Place a known sample weight (minimum 50 g) on the
waste oil scale bracket.
Wait a few seconds then press
ENTER
key to confirm
and wait!
The display shows:
4. Turn
COOLIUS SELECT
to set the weight value.
Press
START
to confirm.
Calibration completed.
5.3 Pressure transducer
IMPORTANT:
The quick coupler must be assembled on the service hoses and the gauges must
show 0 (zero) pressure!
1. Press
START
to begin the calibration of pressure trans-
ducer.
2. The vacuum pump starts and perform a short vacuum
phase.
After a few seconds, the displayed value (ADC points)
must be lower than 250.
Unload used oil
scale!
Confirm by “ENTER”
used oil scale
unloaded!
Place the sample
weight on the scale.
Set the sample
weight value.
g.
Confirm by “START”
Wait!
xx