www.vxitech.com
SVM2608 Programming
37
Sample Rate, High-Speed (0xAC, 0xD4) — Read & Write
D15 – D9
Unused
These bits are reserved for future use.
D8 – D0
SAMPRAT24 – 16
Sample Interval
– These bits set the high speed sample rate.
Bit Weight = 8.333 ns/bit
Minimum Value = 6
Maximum Value = 2
24
– 1
Sample Rate, High-Speed (0xAE, 0xD6) — Read & Write
D15 – D0
SAMPRAT15 – 0
Sample Interval
– These bits set the high speed sample rate.
Bit Weight = 8.333 ns/bit
Minimum Value = 6
Maximum Value = 2
24
– 1
Sample Points (0x10, 0x38, 0x60, 0x88, 0xB0, 0xD8) — Read & Write
D15 – D4
Unused
These bits are reserved for future use.
D3 – D0
SIZE19 – 16
Waveform Capture Size
- In Linear mode, this sets the number of data
points to be captured in a single sweep. In FIFO mode, this sets the
threshold for generating a service request interrupt or status bit. The
maximum size is 1 MSamples (2 Mbytes of data).
Sample Points (0x12, 0x3A, 0x62, 0x8A, 0xB2, 0xDA) — Read & Write
D15 – D0
SIZE15 – 0
Waveform Capture Size
- In Linear mode, this sets the number of data
points to be captured in a single sweep. In FIFO mode, this sets the
threshold for generating a service request interrupt or status bit. The
maximum size is 1 MSamples (2 Mbytes of data).
Pre-Trigger Points (0x14, 0x3C, 0x64, 0x8C, 0xB4, 0xDC) — Read & Write
D15 – D4
Unused
These bits are reserved for future use.
D3 – D0
PRE19 – 16
Size of Pre-Trigger Data
– In Linear mode, this sets the number of data
points to be stored before the trigger occurs. In FIFO mode, this register
is not used.
Minimum Value = 0
Maximum Value = (Capture Size) – 1
Pre-Trigger Points (0x16, 0x3E, 0x66, 0x8E, 0xB6, 0xDE) — Read & Write
D15 – D0
PRE15 – 0
Size of Pre-Trigger Data
– In Linear mode, this sets the number of data
points to be stored before the trigger occurs. In FIFO mode, this register
is not used.
Minimum Value = 0
Maximum Value = (Capture Size) – 1
Trigger Delay (0x18, 0x40, 0x68, 0x90, 0xB8, 0xE0) — Read & Write
D15 – D0
DELAY31 – 16
Trigger Delay
- The trigger delay value is used to set the number of
samples between trigger occurrence and storing of data.
Minimum Value = 0
Maximum Value = 2
32
– 1