![VXI GPIB-VXI User Manual Download Page 24](http://html1.mh-extra.com/html/vxi/gpib-vxi/gpib-vxi_user-manual_1057860024.webp)
Configuration and Startup Procedures
Chapter 2
GPIB-VXI User Manual
2-4
© National Instruments Corporation
Setting the Logical Address
The logical address switches 8 through 1 correspond to the GPIB-VXI's logical address bits 7
through 0, respectively. The
ON
position corresponds to a bit value of 0, and
OFF
corresponds
to a value of 1. For example, to set the logical address of the GPIB-VXI to 25 (19h), set the
switches as shown in Figure 2-2. Notice that setting the logical address to any setting but 0 will
disable the GPIB-VXI RM.
Key
Black = side you must press down
1
2
3
4
5
6
O
F
F
O
N
7
8
Settings for Logical Address = 19h
MSB
Figure 2-2. Example Logical Address Switch Setting
Setting the GPIB Primary Address
The GPIB switches 5 through 1 correspond to GPIB primary address bits 4 through 0,
respectively. The
ON
position corresponds to a bit value of 0, and
OFF
corresponds to a value
of 1. To set the primary address of the GPIB-VXI to 3 (03h), for example, set the switches as
shown in Figure 2-3.
Key
Black = side you must press down
= not used in this context
1
2
3
4
5
6
O
F
F
O
N
7
8
Settings for GPIB Primary Address = 3
Figure 2-3. Example GPIB Primary Address Switch Setting
The primary address switches can be overridden by the nonvolatile memory configuration as
described in the
Change Configuration Information
section of Chapter 4,
Nonvolatile
Configuration
.