 I Instruction manual VWR

®

 Media Pump 

 

 

Programs and calibration 

 

A calibration is necessary for programs distributing doses with a fixed volume. The calibration enables the Media 

Pump to measure the real flow with the conditions of the program, to memorize it and to take it into account at 
each dose delivered. The flow rate being known, the Media Pump makes the number of rotations necessary for 

each dose.  
The flow delivered by the pump will indeed depend on the rotation speed; the given rotation speed of a given 

pump, will depend upon the diameter of the tubing and the viscosity of the liquid. The flow directly proportionally 
depends on the rotation speed but this proportionality depends on the tubing diameter and the viscosity of the 

liquid. 
Moreover, with the same tubing and liquid, the medium flow can vary according to the volume of a dose, 

especially when going from a few microliters to several liters. 
As each program has different parameters: volume of a dose, tubing diameter and liquid viscosity; a calibration is 

necessary for each program. 
 

General settings 

 

The access to programming functions in the general settings is made by the choice of the 

“C O N F I G U R A T I O N ”

 option in the main menu or by pressing the “

change settings” button on the top right of 

the rotating button. 

Here are the options appearing in the settings menu in the general settings: 

 

Flow direction of the pump 

The arrow displayed in front of “

F L O W   D I R E C T I O N ”

 indicates the direction towards which goes the pump 

and consequently the direction in which the tubing must be placed. If the arrow indicates a direction from left to 
right like this:          then the source must be placed on the left side of the pump and the container on the right side.  

To change the direction of the pumping, please select the line and then press the rotating button. Get back to the 
main screen by pressing the “

esc.” button. 

Note: The direction of the pumping is a general parameter, meaning that when using the Media Pump, the 

direction will be set by whatever the program you use. 

Anti-drop system 

The 

Anti-drop option prevents a drop showing up at the end of the tubing at the end of a distribution. Such a 

drop can be a trouble if the user shakes the tubing and the drop falls into the container, because it would alter 

the precision. 
To prevent that a drip appears, the pump goes back and sucks the drop back in the tubing. This operation can 

take a fraction of a second after dispensing. 
To activate or deactivate the option, select the 

“A N T I - D R O P ”

 line and then press the rotating button. Go 

back to the main screen by pressing the “

esc.” button. 

Sound 

Media Pump

 makes a little sound after each operation. 

To activate or deactivate the sound, select the line 

“S O U N D ”

 and then press the rotating button. Go back to 

the main menu by pressing the “

esc.” button. 

C O N F I G U R A T I O N  
 

F L O W   D I R E C T I O N :

Ƈ  

 

A N T I - D R O P :     y e s  
S O U N D :             y e s  
C O N T R A S T  
D A T E   /   T I M E  
L A N G U A G E : e n g l i s h  

Summary of Contents for avantor 710-2126

Page 1: ...Instruction manual VWR Media Pump EU cat no 710 2126 ___ Version 01 Issue 06 2021 For manuals in other languages please visit vwr com...

Page 2: ...nufacturer Europe VWR International bv Researchpark Haasrode 2020 Geldenaaksebaan 464 B 3001 Leuven 3216385011 http be vwr com UK Importer VWR International Ltd Hunter Boulevard Magna Park Lutterworth...

Page 3: ...ttons and Switches 7 Operational keys 9 Getting Started 9 Important Guidelines 10 General settings 11 Operation 12 Calibration 15 Troubleshooting 20 Appendix 20 Repair and maintenance of Media Pump 21...

Page 4: ...ed Media Pump can be used with lab temperature from 5 to 30 C Media Pump is in compliance with ISO 7218 and FDA BAM standards 2 precision can be achieved in many applications Safety Information Avoidi...

Page 5: ...the tube into the peristaltic pump from front to back Be careful to center it to avoid pinching the tubing when closing the pump BE CAREFUL carefully read the tubing installation guide provided When t...

Page 6: ...hell designed to last long Product Specifications Dose volume 50 L to 99 L Dispensing modes Continuous Dose Multi dose Multi dose pause Adjustable from 0 1 to 99 9 s Watson Marlow pump Single pump hea...

Page 7: ...posed of a screen a rotating button and a few other buttons Information and menus appear on screen The rotating button at the right side of the screen is the main control of the Media Pump it allows y...

Page 8: ...n Example of the main menu when the Media Pump is turned on The on screen menu is hierarchical the user can when choosing an option in the menu go to another menu called sub menu For example the optio...

Page 9: ...user controls by hand or by foot the starting moment of the pump for the first dose The pump automatically stops when the programmed volume is dispensed Then it automatically dispenses again after a p...

Page 10: ...apted for dispensing of little volumes On the side of the Media Pump is engraved 1 diagram It shows the precision that you can get according to your volume and the inner diameter of the tubing Example...

Page 11: ...on of the pump The arrow displayed in front of F L O W D I R E C T I O N indicates the direction towards which goes the pump and consequently the direction in which the tubing must be placed If the ar...

Page 12: ...menu choose the option MODIFY PROGRAMS On the screen will appear the list of the programs already created If you scroll the list at the end will appear a list of standard names program 5 program 6 et...

Page 13: ...g Then turn the button to modify the character under the cursor Typographic characters scroll numbers first and then alphabetic letters and punctuation characters Turn the button until the character a...

Page 14: ...ntal mode Launch the program When the pump is working the speed is displayed on screen and you can modify with the rotating button The last chosen figure is memorized in the program Inner diameter of...

Page 15: ...n an optimal precision calibration is recommended Principle of operations The operation consists in pouring a fixed volume of liquid in conditions closest to real conditions to measure it and then ent...

Page 16: ...k possibly by weighing the volume of the poured liquid Calculation of accuracy The accuracy depends on 2 things tubing inner diameter dispensed volume during calibration Tubing inner diameter The tubi...

Page 17: ...d an operation You can temporarily suspend the operation by 3 ways by pressing the START STOP button on the rotating button or on the esc button You can start the dispensing again by pressing the STAR...

Page 18: ...y pressing the START STOP button At the end of all dispensing the screen displays nothing special but that the 2 progression indicators are complete The Media Pump makes a little bip sound if the opti...

Page 19: ...the BASIC PUMPING function in the main menu To do so select the option in the main menu and then press the START STOP button or on the rotating button and then press the same button again Stop the pu...

Page 20: ...the density in the calibration process see chapter Calibration The pump has been opened Calibrate after each opening of the pump Appendix Using the handy gun ref 710 2127 For highly precise work you c...

Page 21: ...the dust and dirt with a soft cloth Once a week grease the pump head with a universal multi purpose grease to the friction areas see diagram below to prevent premature wearing Signs of wearing of pum...

Page 22: ...r for VWR Media Pump 1 710 2130 Set of silicone tubings inner 4 8 mm GL45 screw cap tube weight inox nozzle filter for VWR Media Pump 1 710 2131 Set of silicone tubings inner 6 4 mm GL45 screw cap tub...

Page 23: ...NTY OF MERCHANTABILITY Compliance with local laws and regulations The customer is responsible for applying for and obtaining the necessary regulatory approvals or other authorisations necessary to run...

Page 24: ...R International Ltd VWR International Northern Ireland Ltd Orion Business Campus Northwest Business Park Ballycoolin Dublin 15 Tel 353 01 88 22 222 Email sales ie vwr com Italy VWR International S r l...

Reviews: