Insert the power cord into the power inlet module. Place the retaining
bracket (P/N 195763) over the connector. Tighten retaining screws to
secure.
Note
See the serial tag on the side of the unit for electrical specifications or
refer to the electrical schematics in this manual.
s
The freezer should be operated on a dedicated grounded service. Check the
voltage rating on the serial tag of the unit and compare it with the outlet
voltage. Then, with the power switch turned off, plug the line cord into
the wall outlet.
First, turn on the freezer power switch. Then open the lower front door by
grasping the bottom left corner. Locate the battery switch (Figures 1-3 and
1-4.) and turn it to Standby mode ( ). During initial freezer start-up, the
system battery may require charging and the Low Battery message may
appear in the message center.
Note
Ensure the battery switch is turned to Standby mode ( ). The
rechargeable batteries require 36 hours to charge at initial start-up. A “Low
Battery” alarm may occur until the batteries are fully charged. Should a
power failure occur during the initial start-up period, the electronics will
have limited operation.
s
1-8
ULT Freezer
VWR International
Section 1
Installation and Start-up
Attach Power Cord
Figure 1-11.
Power Cord Retaining Bracket
Connect Unit to Electrical
Power
Summary of Contents for 5700 Series
Page 19: ......
Page 21: ......
Page 41: ......
Page 44: ...ULT Freezer 7 1 VWR International Section 7 Parts ...
Page 45: ...7 2 ULT Freezer VWR International Section 7 Parts ...
Page 46: ...ULT Freezer 7 3 VWR International Section 7 Parts ...
Page 59: ......
Page 60: ...ULT Freezer 8 1 VWR International Section 8 Refrigeration Schematics ULT Freezer 8 1 ...
Page 61: ...8 2 ULT Freezer VWR International Section 8 Refrigeration Schematics ...
Page 62: ...ULT Freezer 8 3 VWR International Section 8 Refrigeration Schematics ...
Page 63: ...8 4 ULT Freezer VWR International Section 8 Refrigeration Schematics ...
Page 64: ...ULT Freezer 8 5 VWR International Section 8 Refrigeration Schematics ...
Page 65: ......
Page 66: ...ULT Freezer 9 1 VWR International Section 9 Electrical Schematics ...
Page 67: ...9 2 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 68: ...ULT Freezer 9 3 VWR International Section 9 Electrical Schematics ...
Page 70: ...ULT Freezer 9 5 VWR International Section 9 Electrical Schematics ...
Page 71: ...9 6 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 72: ...ULT Freezer 9 7 VWR International Section 9 Electrical Schematics ...
Page 73: ...9 8 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 74: ...ULT Freezer 9 9 VWR International Section 9 Electrical Schematics ...
Page 75: ...9 10 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 76: ...ULT Freezer 9 11 VWR International Section 9 Electrical Schematics ...
Page 77: ...9 12 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 78: ...ULT Freezer 9 13 VWR International Section 9 Electrical Schematics ...
Page 79: ...9 14 ULT Freezer VWR International Section 9 Electrical Schematics ...
Page 80: ...ULT Freezer 9 15 VWR International Section 9 Electrical Schematics ...
Page 81: ......
Page 90: ...VWR International P O Box 5015 Bristol CT 06011 8960 ...