Instruction Manual INCU-Line Premium Peltier Cooled
Page
9
/
54
The picture below is an example of the placement of samples in the chamber.
H
m
a
x
H
H
m
a
x
a
u
v
w
x
y
z
b
u+v+w > 1/3 a
x+y+z > 1/3 b
Hmax = 2/3H
u v w
x y z
H
≅
≅ ≅
≅
Front view
Top view
Shelf
Following the above rules will provide optimal parameters of temperature variation and fluctuation.
3.4. Closing the chamber door
The ILPR incubators have been equipped with a 2-point lock. To close the door, turn the handle to the verti-
cal position, push the door gently then turn the handle to the horizontal position. If the door has not been
closed properly, an alarm will sound. Closing the door correctly reduces energy consumption and ensures
proper temperature variation and fluctuation.
3.5. Internal glass door
To open or close the internal glass door use the plastic handle installed on the glass. While operating the
unit at high temperatures, do not touch the glass door or inner parts of the chamber. Always use safety
gloves to protect yourself and minimise the risk of getting burnt.