VWR Collection Manual ver 1, rel 1 20.02.2017
PROBE INSTALLATION
Plug probe sensors into jacks on side of unit, then slide probe channel switch to desired channel.
With probe sensor attached, unit displays current probe temperature & probe MIN/MAX temperatures.
With the probe sensor removed unit displays “--.-”.
PROBES
Model 10050-584 (620-2334)/10050-586 (620-2335)--
Supplied with 1 bottle probe (10050-584 (620-
2334)) or 2 bottle probes (10050-586 (620-2335)), designed for use in refrigerators and freezers.
Bottle probes are filled with a patented nontoxic glycol solution that is GRAS (Generally Recognized
As Safe) by the FDA (Food and Drug Administration) eliminating concerns about incidental contact
with food or drinking water.
The solution filled bottle simulates the temperature of other stored liquids. Velcro
®
and a magnetic
strip are provided to mount the bottle inside a refrigerator/freezer. The included micro-thin probe cable
permits refrigerator/freezer doors to close on it. (Do not immerse bottle probes in liquid).
Model 10050-588 (620-2336) / 10050-590 (620-2337)--
Supplied with standard plastic probe with
cable. Designed for use in air and liquids, the sensor and cable may be completely immersed.
Model 10050-592 (620-2338)–
Supplied with 1 bottle probe with cable and 1 standard plastic probe
with cable.
Model 10050-594 (620-2339) / 10050-596 (620-2340)–
Supplied with 1, 5ml vaccine bottle probe
(10050-594 (620-2339)) or 2, 5ml vaccine bottle probes (10050-596 (620-2340)). Designed to fit in
vaccine trays.
Velcro
®
and magnetic strip are supplied to mount the bottle to the inside of a refrigerator/ freezer and
to mount the display unit to the outside. Micro-cable permits refrigerator doors to close on it.
RESETTING THE UNIT
Use a pointed object to push the RESET button on the back of unit.
NOTE:
HI/LO temperature alarm settings and MIN/MAX memories will be cleared after resetting the
unit.
DISPLAY MODES
The unit has two display modes: MIN/MAX DISPLAY MODE and ALARM DISPLAY MODE. To change
from one mode to the other press the MODE button once.
MIN/MAX DISPLAY MODE - is indicated by the display of current temperature and the MIN
(minimum) and MAX (maximum) temperatures. These values correspond to the probe channel
selected with the probe switch.
ALARM DISPLAY MODE - is indicated by the display of current temperature and the HI and LO alarm
settings. These values corresponds to the probe channel selected by the probe switch.
RESETTING MIN/MAX MEMORIES
There are two ways to reset the memories: