background image

VWR International, LLC., Stereo Zoom Microscope                                     

 

b.  If using the frosted glass plate, insert the blue filter into the centre of the base, with the frosted  

 

surface facing down. 

c.  Place the glass plate on top, again, with the frosted surface facing down. 

d.  Re-tighten the locking set screw. 

 

2.  Before using the stereomicroscope, adjust the light intensity control to minimum position. This  

 

should be repeated every time the stereomicroscope is switched on or off to prolong the use of the  

 

bulb. (Intensity control not available for models with the 2LBB base). 

 

Three light switches are situated on the stereomicroscope base.  

MAIN :

 Turns the entire unit on or off.  

   I  

:

 Turns incident light on or off. (Top illumination) 

   T 

:

 Turns transmitted light on or off. (Bottom illumination) 

 

3.  Press the main switch to position 

“I”

4.  Press the incident or transmitted illumination switch, or both at once, to position 

“I”

 or 

“II”

, according  

 

to necessities in examining the sample. 

5.  Light intensity should be adjusted according to the objectives utilised, or the type of sample  

 

observed.  

6.  The angle of the incident illumination can be adjusted using the screw that permits variation on the  

 

orientation of the illuminator lenses. 

7.  If the stereomicroscope does not work, please check the “Troubleshooting” section of this manual. 

 

 

B.  Interpupillary Adjustment 

1.  Whilst looking through both eyepieces, move eyepiece tubes on their axis, holding head by the  

 

eyepiece tubes housing the prisms. 

2.  When a full field of view is obtained, and converted into a single image, interpupillary distance is  

 

correct for your eyes.   

3.  Each user should adjust interpupillary distance to match their eyesight. 

 

C.  Focusing 

1.  Turn objective turret so that the lowest magnification number 2X is facing the front of the 

microscope,  

 

lined up with the eyepiece, and the 4X magnification is located at the two sides, perpendicular to the  

 

eyepiece. The turret permits a ¼ turn. Ensure that turret is “clicked” into its correct position.    

 

NB:

 For models with N3GG or 2LBB bases, omit steps 2, 3 and 4, passing directly to 5. 

 

2.  Place a flat object, or a specimen slide on the centre of the stage. 

Summary of Contents for 89404-492

Page 1: ...28 Sept 2012 VWR Stereo Microscope Operation Manual Version 1 0 0 VWR Stereo Microscope 89404 492 89404 494 89404 496 89404 498 89404 500 89404 502 89404 492 VWR Stereo Microscope Dual Halogen 2x 4x www vwr com ...

Page 2: ...VWR International LLC Stereo Zoom Microscope 2 ...

Page 3: ...iterature and this manual Before using the product verify that this product is suitable for its intended use Do not modify the system components or use the unauthorized parts as this will void the product warranty The following warning labels or symbols are found on the microscope Study the meaning of the warning labels or symbols and always use the equipment in the safest possible manner Warning ...

Page 4: ...onents and controls can be found in the diagram on page 2 These instructions are based on the assembly of a binocular model model with additional notes applying specifically for the other models in the series For those models with A objectives objectives referred to are 1X and 2X rather than 2X and 4X whilst for models with B objectives objectives referred to are 1X and 3X Unpacking All components...

Page 5: ...lder with extreme care c Re tighten the head locking screw Warning Before connecting the microscope to an electrical source always check that the voltage coincides with that of the stereomicroscope Operation A Starting Up The stereomicroscope comes with two stages One is frosted glass for the observation of microscopic specimens slides or thin or transparent samples such as leaves insect wings and...

Page 6: ...nsity should be adjusted according to the objectives utilised or the type of sample observed 6 The angle of the incident illumination can be adjusted using the screw that permits variation on the orientation of the illuminator lenses 7 If the stereomicroscope does not work please check the Troubleshooting section of this manual B Interpupillary Adjustment 1 Whilst looking through both eyepieces mo...

Page 7: ...left hand eyepiece tube there is a diopter adjustment collar Its normal position is when the lower part of the collar is aligned with the sign marked on the eyepiece tube In the case of differences in eyesight 1 With the right eye look through the right eyepiece and adjust focus 2 Now use the left eye to look through the left eyepiece and adjust the focus by turning the diopter adjuster on the lef...

Page 8: ...RE ATTEMPTING ANY MAINTENANCE PROCEDURE TO AVOID THE RISK OF ELECTROCUTION CONSULT YOUR DISTRIBUTOR IF ANY REPAIR OR MAINTENANCE PROCEDURE IS REQUIRED TO YOUR MICROSCOPE THAT DOES NOT APPEAR IN THIS INSTRUCTION MANUAL A Optical maintenance Do not attempt to disassemble any optical component For any repair work not specified in this manual consult the technical service responsible in your area Befo...

Page 9: ...ew down firmly Changing incident illumination bulb a Unscrew the illuminator protector tube turning it anti clockwise and removing the tube from that of the lenses b With a cloth carefully pull out the bulb to disconnect it from the socket c Do not touch the new bulb with bare hands use a clean cloth Insert the pins of the bulb in the socket d If the bulb is accidentally touched with bare hands it...

Page 10: ...LUTION The bulb does not work Outlet inoperative Cable not connected Bulb burned out Fuse blown Wrong Bulb Have it repaired by a qualified specialised technician Connect the cable to the power source Replace bulb Replace fuse Replace with appropriate bulb Bulb has short life span Very high voltage Reduce light intensity to the minimum before turning on or off the stereomicroscope Bulb burns out im...

Page 11: ... may also be attributed to dirt on the inside of the eyepiece It is recommended therefore that the lens be cleaned by a recognised service technician Mechanical Problems PROBLEM CAUSE SOLUTION It does not stay in focus The head drops down Adjust the tension of the coarse focus knob Moving the microscope If possible avoid moving the stereomicroscope Carry the stereomicroscope in both hands One hand...

Reviews: