CONTROL PANEL OVERVIEW
(See Figure One)
5.1 Power
Switch: The main power switch on the control panel (green lighted I/O)
controls all power to the oven. It must be in the I/On position before any systems
are operational. The green pilot light in the switch will be lighted when the switch is
in the ON position.
5.2 Timer
Switch: The black I/O power switch marked TIMER is located to the right of
the main power switch. It controls the power to the time circuit. In the O/Off position
the oven heat is controlled with no timed duration. In the I/On position heat is
controlled for a timed interval and then the heat shuts off.
5.3
Overtemperature Safety Thermostat: This control is marked HIGH LIMIT and is
equipped with an adjustment knob and a graduated dial from 0 – 10. It is
independent of the Main Controller and guards against any failure which would
allow temperature to rise past the main controller’s set point. This allows continued
operation of the oven until the problem can be corrected or service can be arranged.
It is not recommended that the unit be operated for extended periods of time using
only the Overtemperature Safety as the temperature controller. On “S” suffix
models, the Overtemperature Safety will shut the oven off by tripping the over-
temperature relay. Pushing the red reset button is required to restart the oven after
the interruption reason has been cleared.
5.4 OTP
Light: This pilot lamp marked OTP is directly above the Overtemperature
Safety Thermostat. The light will come on when the Safety Thermostat has been
activated and taken over control of the oven. The power to the over-temperature
relay coil is turned OFF for “S” suffix models. Under normal operating conditions
the pilot lamp should never be lit.
5.5
Manual Reset Button: (“S” suffix models only) The red reset button, next to the
OTP light, is for resetting the oven power if the temperature relay has been de-
activated. The reset button must be pushed for initial start up and any time
thereafter when there is a power interruption to the Overtemperature relay.
5.6 Timer
Control: This control is marked SET/TIMER and consists of a digital display,
UP/DOWN arrow pads, a RESET “PUSH” pad, a START/STOP “PUSH” pad and a
TIMER ACTIVATED light. This control provides the ability to set a timed heat
interval, activate the start-up of the timed heat cycle and shut down the timed heat
cycle automatically.