VTI Instruments Corp.
78
EX1401 Instrument Drivers
Alarms
LXI alarms are a mechanism that generates events at a fixed time interval as specified by IVI 3.15,
IVILxiSync specification. When the
Enabled
property is true, the alarm starts at the time specified
by
TimeSeconds
and
TimeFraction
and repeats for
RepeatCount
times. The
Period
property defines
the interval between the two successive alarm events. Setting the
RepeatCount
property to 0 causes
the alarm to continue firing at the specified
Period
until the
Enabled
property is set to false. Setting
the
Period
to 0 causes the alarm to fire only once. The alarm properties can only be modified when
the alarm is disabled.
As an extension to the LXI alarm capabilities, the EX1401 also supports “now” alarms (similar to
“now” LAN events). Setting the
TimeSeconds
and
TimeFraction
properties to 0 causes the alarm to
fire as soon as it is enabled.
Alarms can be used as an arm or a trigger event source.
Summary of Contents for EX1401
Page 2: ......
Page 24: ......
Page 57: ...www vtiinstruments com EX1401 Web Page Operation 57 ...
Page 94: ......
Page 98: ......
Page 102: ......
Page 111: ...www vtiinstruments com EX10xxA Index 111 FIGURE A 1 IEEE1588 NETWORK TOPOLOGY ...
Page 112: ......