background image

Operating instructions, Quick installation  

Power supply  

Terminals 17, 18.  
The AST 3 module is powered by 24 V DC.  
Rail mounted power supplies can be ordered  
from Nobel Weighing Systems.  

Analogue output  

Terminals 24, 25.  
The measurement value is presented as a current  
or voltage signal, according to set-up selection.  
Connect the cable shield to ground, preferably to  
a ground terminal at the mounting rail.  

Serial communication  

Terminals 19 – 23.  
Set-up by serial communication is needed to take  
AST 3B into operation. All AST 3 have a serial port  
for RS-485 on 2-wires or 4-wires with common  
earth (COM). If necessary, a converter unit should  
be connected to adapt the interfaces.  

The line must have 120 ohm termination at both  
ends. For AST 3 it should be mounted on separate 
terminals and connected according to the diagrams. 
Refer to manufacturer instructions for termination  
at the control unit (PC), or at the converter unit.  

Connect the cable shield to ground, preferably to  
a ground terminal at the mounting rail.  

Set-up programme deltaCOM  

AST 3B must be connected to transducer(s) and power supply, and also to a 
PC with Windows 95/98/2000/XP/NT4.0 to make set-up possible. Included in 
the AST 3 delivery is a diskette with the set-up programme deltaCOM. This 
diskette also contains instructions for connection and set-up of the serial 
communication between the instrument and the PC.  

To install the programme, insert the diskette in the computer, press the ‘Start’ 
button and select ‘Run...’ from the menu, type [drive letter]:\setup in the dialogue 
box and then follow instructions on the screen.  
If nothing else is stated, the programme deltaCOM and a ReadMe file will be 
found in Start > Program > deltaCOM.  

By use of deltaCOM it is possible to find all AST 3 modules communicating with  
the PC, and to study and edit all set-up parameters for these modules.  

IN1

IN2

15

16

+

OUT+

OUT–

24

25

2-wire

*

Terminating resistor on separate
terminals at the last unit on the line.

120Ω

19

20

21

22

23

TXD–

TXD+

RXD–

RXD+

COM.

4-wire

Terminating resistor on separate 
terminals at the last unit on the line.

*

*

19

20

21

22

23

TXD–

TXD+

RXD–

RXD+

COM.

120Ω

*

120Ω

*

Summary of Contents for BHL Nobel AST 3

Page 1: ......

Page 2: ......

Page 3: ...supply 2 Start up 2 View alternatives 3 Zero setting 4 Installation Mechanical installation 5 Electrical installation 5 Set up programme deltaCOM 6 Quick set up calibration General 7 Common parameter...

Page 4: ......

Page 5: ...nt Additional installation and set up of more instrument functions not covered by this description can also be performed For a complete instrument description refer to publication 600 232 AST 3 Techni...

Page 6: ...d for about five seconds Then AST 3P is automatically switched over in Operation mode If the instrument is set for operator start up the text Press ENTER to start AST is displayed until the key to the...

Page 7: ...tual analogue output signal Level settings This view displays the switching levels for the two supervising channels If these channels are not in use level 0 is displayed Input signal This view display...

Page 8: ...enu Quick set up is displayed Press ENTER 5 The first parameter name Language is displayed Press a number of times until Set zero is displayed The parameter value for Set zero is a live weight value 6...

Page 9: ...les with other frequencies or high power an even wider distance is preferable 4 wire connection should be used if the integrated transducer cable can be connected directly at AST 3 By 4 wire connectio...

Page 10: ...preferably to a ground terminal at the mounting rail Set up programme deltaCOM AST 3B must be connected to transducer s and power supply and also to a PC with Windows 95 98 2000 XP NT4 0 to make set...

Page 11: ...e most accurate calibration method where known weights are used to give well defined loads on the scale Before selection of calibration method the value of some common instrument parameters must be se...

Page 12: ...ess to step forwards to next alternative or press to step backwards to the previous alternative until the correct alternative is displayed Language alternatives Press ENTER for 2 seconds to accept the...

Page 13: ...i e the weight value that corresponds to full range at the analogue output Press to get the parameter Capacity displayed Capacity 500 00 kg Press ENTER to make editing possible Capacity 0500 00 kg A...

Page 14: ...d by a cursor at the parameter value Then one of the alternatives must be chosen and accepted Press to get the parameter Calibration type displayed Calibration type Data sheet The value points out whi...

Page 15: ...eet unit is Newton N and the measurement unit is kg If data sheet unit and measurement unit are equal the value should be 1 00000 Press to get the parameter Conv factor displayed Conv factor 9 80665 P...

Page 16: ...the rated output should be set to 0 00000 Press to get the parameter Rated output 1 displayed Rated output 1 2 03900 mV V Press if the parameter value should be edited and make the editing like in po...

Page 17: ...r values should be saved i e copied to a specific memory in the instrument Alternatively the new values can be cancelled and all parameters will resume the values they had before Quick set up was star...

Page 18: ...isplayed parameter value is the load on the scale that was set for the lower calibration point in the last calibration normally zero unloaded scale Press ENTER Value cal p 1 XXXX XXX kg Actual measure...

Page 19: ...n to make editing possible Value cal p 2 0500 00 kg The displayed parameter value is the value for calibration point 2 from the last calibration with leading sign and a cursor The parameter value can...

Page 20: ...Set zero displayed and check that the scale is unloaded Set zero XXXX XXX kg Press ENTER to set the parameter value to zero Set zero 0000 00 kg The value is set to zero with leading sign and a cursor...

Page 21: ...ck set up was started Press ENTER to get the sub menu displayed Save changes No Esc Yes Press if you do not wish to exit from the set up mode Press to answer Yes and save the new parameter values Pres...

Page 22: ...lue cal p 2 500 0 kg 2 03900 mV V 2 03900 mV V 2 03900 mV V Weight value 123 4 kg 001 R1 On R2 On Start up Reset Quick set up Quick set up Language Resolution Measurement unit Main menu English kg 0 1...

Page 23: ...rement unit kg Resolution 0 1 Capacity 500 0 Ana output type 4 20mA Calibration type Data sheet Conv factor 9 80665 Number of transd 3 Rated load 2000 0 Rated output 1 2 03900 Rated output 2 2 03900 R...

Page 24: ...Operating instructions Quick installation...

Page 25: ......

Page 26: ...Document no 35158 Article no 600 243 R11 Vishay Nobel AB 2011 05 11 Subject to changes without notice www blhnobel com...

Reviews: