8 | Operation
KBM-110
24 / 32
B-H86.0.12.DB214-1.0
Setting the gradient
PK.SL
Current measure-
ment
Setting of the gradient for calibration of the pH
measurement. If a calibration cannot be carried out,
continue with the
function key
Input
INP
PK
0RP
mV
Redox in mV, relative to silver / silver chloride -
electrode
0RP
mVH
Redox in mV
H
, relative to hydrogen electrode
Temperature unit
UN,T
°
(
Temperature display in °C
°
F
Temperature display in °F
Alarms
AL.
OFF
No active alarm
ON
Alarm alerting via text display, acoustic signal and
flashing of the backlighting
BEEP
Alarm alerting via text display and acoustic signal
L,TE
Alarm alerting via text display and flashing of the
backlighting
AL.LO
Depending on the setting of the parameter value
INP
PK
0.00 ..
AL.K,
mV
-1500 ..
AL.K,
mV
H
-1293 ..
AL.K,
AL.K,
Depending on the setting of the parameter value
INP
PK
AL.LO
.. 14.00
mV
AL.LO
.. 1500
mV
H
AL.LO
.. 1707
Shut-off time
POFF
OFF
No automatic shut-off
15 30 60 120 240
Automatic shut-off after a selected time in minutes,
during which no buttons have been pressed