50
a) Measuring tolerances
The accuracy is valid for one year at a temperature of +23 to +25 °C (+73 to +77 °F), and at a relative humidity of
less than 75%, non-condensing.
1
Infrared temperature measuring area
Measuring area °C
Accuracy
Reproducibility
-50 to +20 °C
±5 °C
±2 °C
+20 to +500 °C
±(1.5 % +1.5 °C)
±1 % or ±1 °C
+500 to +1000 °C
±2 %
±1 % or ±1 °C
+1000 to +1600 °C
±2.8 %
±1.5 %
Measuring range °F
Accuracy
Reproducibility
-58 to +68 °F
±9 °F
±4 °F
+68 to +932 °F
±(1.5 % + 3 °F)
±1 % or ±1.8 °F
+932 to +1832 °F
±2 %
±1 % or ±1.8 °F
+1832 to +2912 °F
±2.8 %
±1.5 %
2
Contact temperature measurement type K
Measuring area °C
Accuracy
-50 to 0 °C
±3.5 °C
0 to +1370 °C
±(1 % + 2 °C)
Measuring range °F
Accuracy
-58 to +32 °F
±7 °F
+32 to +2498 °F
±(1 % + 4 °F)
3
Air temperature measurement
Measuring area °C
Accuracy
0 to +50 °C
±1 °C
+10 to +40 °C
±1.5 °C
Residual range
Measuring range °F
Accuracy
+32 to +122 °F
±1.8 °F
+50 to +104 °F
±3 °F
Residual range
Summary of Contents for 1274960