background image

D-11049-2, Ver 11 

 

Page 20 of 26 

 

 WARNING 

Hazard:

 The patient experiences discomfort when the ear tip is inserted into the ear canal. 

Remedy:

 

The operator must check whether the ear tip selection is incorrect and replace with a properly fitting ear tip. 

Hazard:

  Audible levels are uncomfortable for the patient. 

Remedy:

 

The operator must immediately select a lower setting for the stimulus levels. 

Hazard:

 In ABR testing skin at the site of the electrode application is damaged, for example, irritated (red), scratched, bruised, sore, cut, wounded, 

or bleeding. 
Remedy:  Do  not  conduct  skin  preparation  and  electrode  application  on  areas  of  damaged  skin.  Choose  another  electrode  location,  wait  until 
healing is complete, consult a dermatologist, or refer the patient to a healthcare professional trained to deal with skin damage. 
 

 WARNING 

Electronic Disposal 

 

DO NOT dispose of equipment in unsorted, municipal garbage. 

 

Take discarded equipment to the appropriate electronic waste disposal facility. 
 

Environmental Conditions for Transport and Storage 

Integrity (with Battery Pack) 

Integrity (without Battery Pack) 

 

Ambient temperature of -20 

C to +50 

C (for 30 days) 

 

Ambient temperature of -20 

C to +40 

C (for 90 days) 

 

Ambient temperature of -20 

C to +30 

C (for 1 year) 

 

Relative humidity of 10% to 100% 

 

Atmospheric pressure range of 500 hPa to 1060 hPa 

 

Ambient temperature of -40 

C to +70 

 

Relative humidity of 10% to 100% 

 

Atmospheric pressure range of 500 hPa to 1060 hPa

 

 

Environmental Conditions for Use 

 

Ambient temperature of 10 

C to 39 

 

Relative humidity of 30 % to 75 % 

 

Atmospheric pressure range of 700 hPa to 1060 hPa 

 

 CAUTION 

Following  the  "Guidance  for  Industry,  FDA  Reviewers  and  Compliance  on  Off-The-Shelf  Software  Use  in  Medical  Device",  issued  by  FDA  on 
September  9,  1999,  the  customers  should  not  use  any  software  other  than  that  specified,  as  it  will  violate  the  safety,  effectiveness  and  design 
controls of our Integrity V500 and that such use may result in an increased risk to users and patients. 

Unauthorized  alteration  of  software  provided,  including  downloading  non-validated  or  unauthorized  off-the-shelf  software  may  void  or  breach 
existing service agreements and warranties. For additional information, contact Vivosonic Inc.'s Customer Support. 

 

 CAUTION 

Please follow these recommendations to ensure the test data is accurate. 

 

Substitution of any supplied components with different components may result in measurement error. 

 

No other software may be installed onto the computer on which the Integrity™ software is installed. 

 

Handle the system with care. 

 

The system requires calibration of the transducers – A81/A82 Amplitrodes, A90 VivoAmp, Insert Earphones, B71W Bone Conductor, OAE 
probes,  and  EP  headphones  –  annually  on  each  anniversary  from  the  date  of  manufacture,  and  every  time  a  transducer  is  dropped, 
subjected to mechanical shock, or immersed in a liquid substance. Otherwise, stimuli presented to the patient as specified may lead to 
incorrect test results and misdiagnoses. 

 

DO NOT force the ear tip into the ear canal. 

 

Always carry this device with you when traveling to avoid mishandling and damage of stored or checked luggage. 

 

DO  NOT  ship  this  device  for  service  in  packaging  other  than  the  packaging  supplied  with  the  system  from  the  manufacturer,  or 
comparable packaging. 

 

Signal inputs and outputs are intended for analysis only in connection to the specified equipment described in this manual. 

 

The product DOES NOT require sterilization. Alcohol can be used to clean the transducers. 

 

This equipment does not have protection against water penetration (IPX0). 

 

The system can only be repaired by Vivosonic Inc. or authorized dealers of Vivosonic Inc. 

Summary of Contents for Integrity V500

Page 1: ...Generation 2 G2 Please refer to the Integrity User s Manual D 11049 for complete software operating instructions A PDF of the User s Manual is on the desktop of your Integrity computer QUICK REFERENCE ...

Page 2: ...grity V500 system They have been tested and verified to work with our system and provide the optimal results To order additional electrodes please contact Vivosonic Customer Support at 877 255 7685 Canada US or 1 416 231 9997 Intl WINDOWS INTEGRITY V500 SETTINGS Your Integrity V500 system has been optimized to work with the Windows Operating System Vivosonic has configured the system settings to w...

Page 3: ...plitrode is protected by US Pat No 7 206 625 and 7 548 774 Other patents pending in the US and other countries Integrity Amplitrode VivoAmp VivoLink and Vivosonic are all either registered trademarks or trademarks of Vivosonic Inc in the U S and other countries All other trademarks are the property of their respective owners COPYRIGHT NOTICE 2017 This Quick reference guide is copyrighted by Vivoso...

Page 4: ...ould be taken if the incident occurs CAUTION Messages with this heading indicate information regarding any special care to be exercised by the practitioner and or patient for the safe and effective use of the device All precautions should be followed to ensure data and system integrity ATTENTION Messages with this heading indicate a possible loss of data Follow the procedures to ensure data integr...

Page 5: ...before use Clean surfaces of equipment with disinfectant Eliminate where possible the electro magnetic noise in the room INSTALL THE BATTERY PACK INTO THE VIVOLINK Battery packs have a life expectancy of 5 7 months You may have to replace the battery packs after this period 1 Open the VivoLink battery compartment 2 Remove and set aside any previously used batteries 3 Place the battery pack in the ...

Page 6: ...ed from the list by clicking on the left most row header shown the grey area shown below 6 Click on the Test button to display the Test Window 7 Select a Test Type from the side menu 8 Wait until the initialization process is complete This may require up to 20 to 30 seconds The wireless Bluetooth connection indicator blinks on and off during this time 9 When initialization is completed check the f...

Page 7: ...imited shelf life after which they start to lose their effectiveness SKIN ELECTRODE AND AMPLITRODE VIVOAMP CLIP APPLICATION After proper skin preparation procedures electrodes can be attached to the skin The electrodes are intended for single use only and should not be reused to prevent patient cross contamination and to ensure good electrode skin contact To apply an Ambu Neuroline Electrode to th...

Page 8: ...s use the spring release button on the VivoAmp clips when clipping to or unclipping from skin electrodes do not pull on the VivoTab electrodes EAR TIP INSERTION Before starting a test using Insert Earphones ear tips the ear canal must be clean free of damage or disease and clear of any obstructions to ensure good continuity between the surface of the ear tip and the tissue of the ear 1 Inspect the...

Page 9: ...nes Use when performing ABR testing using air conduction B71W Bone Conductor Use when performing ABR bone conduction testing To perform data collection connect the B71W Bone Conductor to either the left or right mastoid or the forehead Secure the headband supplied with the B71W Bone Conductor Ensure that the entire circular surface of the Bone Conductor is in full contact with the skin Only use th...

Page 10: ...will be displayed stating either that the level is acceptable or it will indicate which electrode needs to be reapplied Before reapplying the electrode repeat the skin preparation procedure See User s Manual on Computer Desktop RUN THE TEST 1 Select the appropriate protocol Please refer to the Integrity User s Manual PDF on the computer desktop for more detailed instruction 2 Select the appropriat...

Page 11: ...OOSE A TRANSDUCER PREPARE THE PATIENT Refer to the Patient Preparation section for Electrode application Ear preparation and ear tip insertion Skin preparation ELECTRODE SITES FOR ABR SCREENING High forehead black clip Shoulder green clip ground Nape of neck blue clip RUN A SCREENING TEST 1 Select the Auto ABR Test button on the Main screen 2 Enter information into the Patient Information screen o...

Page 12: ...ested to display the EcochG trace AP up 9 Clip the non inverting Amplitrode or 1 2 VivoAmp clip to the snap on the cable that corresponds to the other ear Use the reverse set up to get the EcochG trace AP down APPLY GROUND ELECTRODE AND CLIP FOR 1 CHANNEL ECOCHG A81 A90 For EcochG testing one snap electrode is applied to the low forehead of your patient This electrode functions as the ground elect...

Page 13: ... be placed in a chair or on a couch whichever is more comfortable for the patient and the operator Tests are conducted with the patient relaxed lying down or in a chair with the neck supported Patients are encouraged to sleep during the test Infant Patient Tests are conducted with the infant sleeping or lying down and very relaxed Place the VivoLink next to the infant close enough for the Amplitro...

Page 14: ...sure that your OAE probe is correctly calibrated 1 Place your OAE probe into the OAE probe holder 2 Select the Cavity Check button on the Test Window 3 Wait for the results to appear in the message area of the Test Window 4 You can proceed with testing when the messages indicate Cavity Check Passed If cavity check fails inspect the probe for debris and clean the probe if necessary Reinsert the OAE...

Page 15: ... couch whichever is more comfortable for the patient and the operator Tests are conducted with the patient relaxed lying down or in a chair with the neck supported Patients should remain awake Infant Patient 40 Hz ERP testing may not yield repeatable results in infants If nevertheless you wish to conduct tests on infants conduct the tests with the infant lying down and very relaxed but not while t...

Page 16: ...se sensitive It is set in the System Window The default password is blank 3 Highlight a record in the database by clicking on the leftmost row header grey area to preview the test results Each test type has a different database preview window seen in the figures below QUERY THE DATABASE 1 Select a column in the query area of the Database Window For example Patient Info is a common column to query ...

Page 17: ...ard drive failure it is recommended that your database be backed up regularly The system backs up automatically as well when the system is closed BACKUP YOUR DATABASE 1 Select the System button to display the System Window 2 Select the Backup Database button 3 Complete the Save Backup File As window with the location and name of your BAK backup file RESTORE YOUR DATABASE FROM A PREVIOUS BACKUP 1 S...

Page 18: ...ther batteries may damage the charging system and or the batteries and endanger the user Use only the Battery Charger REF 100351 and Battery Pack Cradle REF 100410 supplied with your product to charge the Vivosonic Battery Packs REF 100020 Do not use the Battery Charger REF 100351 and Battery Pack Cradle REF 100410 to charge any other batteries The Battery Packs are meant to be replaced after abou...

Page 19: ...uipment is not suitable for use in the presence of an inflammable anesthetic mixture with air oxygen or nitrous oxide Changes or modifications not expressly approved by Vivosonic Inc could void the user s authority to operate the equipment To avoid accidental swallowing of inhalation of an ear tip do not leave ear tip s within reach of a child Patients should be attended during preparation and tes...

Page 20: ... by FDA on September 9 1999 the customers should not use any software other than that specified as it will violate the safety effectiveness and design controls of our Integrity V500 and that such use may result in an increased risk to users and patients Unauthorized alteration of software provided including downloading non validated or unauthorized off the shelf software may void or breach existin...

Page 21: ...hetic the r h should be at least 30 EFT IEC 61000 4 4 2kV Mains 1kV I Os N A Battery Operated Device N A Surge IEC 61000 4 5 1kV Differential 2kV Common N A Battery Operated Device N A Voltage Dips Dropout IEC 61000 4 11 95 Dip for 0 5 Cycle 60 Dip for 5 cycles 30 Dip for 25 Cycles 95 Dip for 5 Sec N A Battery Operated Device N A Power Frequency 50 60 Hz Magnetic Field IEC 61000 4 8 3A m 3A m Powe...

Page 22: ... the following two conditions 1 this device may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation of this device CAUTION Exposure to Radio Frequency Radiation This device must not be co located or operating in conjunction with any other antenna or transmitter WARNING Changes or modifications not expressl...

Page 23: ...emove particles from the sound canals and vent with the wax removal tool The vent a tiny hole above the cable at the back of the Probe is to ensure the release excessive pressure from the occluded ear canal DISINFECTING THE OAE PROBE 1 The Probe should be wiped with disinfecting wipes after each use Hold the Probe tip down to avoid any ear wax or sanitizing liquid from entering the microphone soun...

Page 24: ...red H 800 1 Item if ABR Screening is Ordered REF Model Part Qty VivoAmp A90 1 VivoAmp Alligator clips WA065 1 Insert Earphone ER 3C 800 1 VivoTab Electrodes 100001 10 pkgs Insert Earphones Eartips Kit RA063 1 Ear Phone Adapter Set ER3 06 ABR 1 Foam Eartips 10 mm ER3 14B Manufacturer Etymōtic Research 1 pkg Silicone Eartips 3 5 mm ER3 14D Manufacturer Etymōtic Research 1 pkg PDI Electrode Prep Pads...

Page 25: ...njury or damage to health is possible IEC 62304 2006 Wireless Communication Type Bluetooth 2 402 to 2 480 MHz Hopping Contains transmitter module FCC ID ED9LMX9838 IC 1520A LMX9838 Wireless Signal Strength 0 66 mW EIRP Emission Designation 2M00G1D Duty Cycle 77 02 Wireless Communication Range 30 feet 10 meters Dimensions of Main Unit L 7 1 18 cm x W 3 6 9 1 cm x H 1 2 3 2 cm Weight of Main Unit 0 ...

Page 26: ...e Cell battery AA Cell size NiMH Nickel Metal Hydride battery NiMH CE Mark with notified body number TÜV SÜD certification 81763 FCC mark The National Institute of Metrology Standardization and Industrial Quality Inmetro UL do Brasil UL of Brazil 02864 09 05756 Agência Nacional de Telecomunicações Brazilian telecommunications regulator our registration number The equipment shall not be disposed of...

Reviews: