background image

 

 

AEN0013-01 Vivago FIDO™ User Manual 

 

Radio transmission 

Range 

Up to 100 meters in free air, 15-20 meters in normal buildings. 

NOTICE:

    Massive  concrete  and  metal  constructions  hinder 

transmission of the RF-signal 

RF frequency range 

868.95 MHz 

Effective radiated power 

10 mW Max 

Operating conditions 

Storage/ transportation 
temperature 

10 – 40 

°

Operating temperature 

10 – 40 

°

Waterproof 

IPx7 

Storage/ transportation pressure  500 – 1100mbar 

Operating pressure 

500 – 1100mbar 

Storage/ transportation humidity  0–99% RH (IPx7) 

Operating humidity 

0–99% RH (IPx7) 

Lights 

Alarm LED 

Red LED, indicates for sending Manual ALARM and battery status 

Buttons 

Buttons 

Vivago (MANUAL ALARM) 

Required power for triggering 
Manual ALARM 

~ 6 N  

Required time for triggering 
Manual ALARM 

~ 2 sec 

 

Usage environment 

The  device  complies  with  the  applicable  EMC  requirements.  The  electromagnetic  field  caused  by  the 
device is weak and not likely to cause interference in nearby electronic equipment.  

Use  of  this  instrument  in  a  dry  environment,  especially  if  synthetic  materials  are  present  (synthetic 
clothing, carpets etc.) may cause damaging static discharges that may cause erroneous alarm situations. 

 

 

Summary of Contents for FIDO

Page 1: ...Vivago FIDO User Manual AEN0013 01 2020 03 07...

Page 2: ...ackage 4 4 Warnings and precautions 4 5 Using the device 5 5 1 Buttons and indicators 5 5 2 Preparations for use 6 5 3 Daily use 6 5 4 Cleaning the device 6 6 Charging the device 7 7 Technical specifi...

Page 3: ...is user manual No ALARMs are delivered during breaks in communication or connections 2 Clarification of the symbols The following symbols may appear in product product packing or its parts Consult ins...

Page 4: ...he user manual are not followed the device may not function correctly Do not use or store the device in hot environments for example in the sauna Do not charge the device battery in hot environments T...

Page 5: ...Alarm LED is a red indicator light located at the upper left corner of the device top cover The Alarm LED indications Constant ON o When not in charger ALARM sent Turns ON when ALARM is sent and remai...

Page 6: ...each of the base station radio signal to ensure proper communication connection Notice that Functional trouble in telephone lines or other telecommunication links may also hinder the delivery of the A...

Page 7: ...s soon as possible but at the latest within three days from the start of the low charge condition to make sure that the device works properly To charge the device place the charger on it as shown on t...

Page 8: ...Mode of operation Continuous automatic Alive messages sent every 25 minutes Protection against electric shock Class II The mains plug of the charger is used as the isolation means from the supply mai...

Page 9: ...sportation humidity 0 99 RH IPx7 Operating humidity 0 99 RH IPx7 Lights Alarm LED Red LED indicates for sending Manual ALARM and battery status Buttons Buttons Vivago MANUAL ALARM Required power for t...

Page 10: ...structions nor by an external factor outside the manufacturer s reasonable control or normal wear and tear Under no circumstances is the manufacturer liable for any indirect damages caused by the syst...

Page 11: ...vidual care Wellbeing information can also be shared with relatives As a pioneer in health care technology Vivago already ensures safety for tens of thousands of individuals in Europe Vivago s unique...

Reviews: