10
± 0.2°C (± 0.4°F): 35.0°C - 42.0°C (95°F - 107.6°F)
± 0.3°C (± 0.5°F): 42.1°C - 43.0°C (107.7°F - 109.4°F)
Object mode:
± 4°C (± 7.2°F): 0°C - 4.9°C (32°F - 40.8°F)
± 1°C (± 2°F): 5.0°C - 60.0°C (41°F - 140.0°F)
± 4°C (± 7.2°F): 60.1°C - 100.0°C (140.1°F - 199.9°F)
Display resolution
0.1°C / 0.1 °F
Latency Time
1 second
Abnormal state display
LCD displays “L °C” if the measured temperature is below the minimum of
measurement range.
LCD displays “H °C” if the measured temperature is over the maximum of
measurement range.
LCD displays “Er1” if the temperature measurement Circuit is abnormal (fault of
SENSOR or the temperature measurement circuit), or if the calibration process
is not completed or EEPROM is abnormal. LCD displays “H” or “L”, if the
ambient temperature exceed 10°C - 40°C (50°F -104°F).
Sound
Volume
≥
50db (the perpendicular distance from db volume sensor to
thermometer is 10cm)
Automatic Shutdown
function
About 60s
Low-voltage Display
function
The product shall display low-voltage signal if the Voltage is below 2.6V ± 0.1
Summary of Contents for VivaGuard EF-100A