15. Technical Data / Ordering Data
15.1 Items supplied
Digital Buret VITLAB continuous E / RS
with GL 45 connecting thread and threaded adapter made of PP in sizes GL 32, GL 38 and S 40 (buttres
thread), telescoping filling tube (200 - 350 mm), telescoping discharge tube (140 - 220 mm), 2 Micro
batteries 1,5 V (LR03, AAA), performance certificate and this operating manual.
Calibration conditions
Distilled water,
delivered at
20 ± 0.5 °C
Number of
testing procedures
10 according to
DIN EN ISO 8655/6
(Technical specifications subject to change.)
15.2 Power and current data
- Operating voltage 3 V
(2 LR03/AAA,1.5 V batteries each)
- Power consumption less than10 mA
- Protection type: IP54
Volume per turn**
Systematic error*, A%
Random error*, CV%
Cat. No.
continuous E
2,5 ml
<_
± 0,2
<_ 0,1
1620506
continuous RS
5,0 ml
<_
± 0,2
<_ 0,1
1620507
* Values based on nominal volume
** Dispensing volume per rotation of the hand wheels
Threaded bottles, coated
Volume
Thread
Cat. No.
1000 ml
GL 45
1671500
2500 ml
GL 45
1671510
15.3 Accessories and Spare Parts
1. Telescopic intake tube (11)
(200 - 350 mm)
(FEP, ETFE, PTFE)
Cat.No. 1671085
2. Discharge/recirculation valve (13)
(
PTFE, PFA)
for continuous E and RS
Cat.No. 1655085
3. Discharge tube (14/15),
complete
Cat.No. 1650135
4. Discharge tube support (16),
adjustable
Cat.No. 1650162
5. Drying tube,
complete
(without drying agent)
Cat.No. 1671095
6. Micro batteries 1,5 V
(LR03/AAA),
2 units
Cat.No. 1670216
7. Plastic stand, PP
Cat.No. 1671116
Threaded adapter
Thread
Cat. No.
GL 45 - GL 32
1670180
GL 45 - GL 38
1670110
GL 45 - S 40
1670120
GL 32 - NS 19/26
1670066
GL 32 - NS 24/29
1670067
GL 32 - NS 29/32
1670068
32