ENG
21
Page
of 36
09260 Issue 1
Vitalograph VitaloJAK - Instructions for Use
Possible
Solutions: (In
probable order)
• The Real Time Clock was not set/reset. If
recording, shut down the device using the
key. If not recording, press Select (forward
arrow key) to resume normal operation.
Reconfigure the date/time on the device.
Problem Fault
Symptoms:
• Buzzer beeps loudly and warning displays
remaining recording time
Possible
Solutions: (In
probable order)
• Occurs when the key is used to turn the
device off during a recording.
• Turn the key so the power switch is back in
the ‘ON’ position to resume the recording.
• Press Select (forward keypad arrow/mute
button) to end the recording and shut
down the device. If the device is left for 10
seconds, it will shut down automatically
and the recording will immediately end.
Problem Fault
Symptoms:
•
Device freezes during configuration or
data upload
Possible
Solutions: (In
probable order)
• Ensure battery pack is disconnected
from the device when connected to the
Vitalograph Web portal.
•
To continue configuration, turn the device
off and on again.
• To restart data upload, disconnect then
reconnect USB data cable
•
Reconfigure or restart data upload
Note:
Contact details are listed at the start of this manual for
assistance with set up, use or maintenance of equipment or to
report unexpected operation or events concerning your Vitalograph
device.
Warning: No modification of this equipment is allowed. Any
unauthorised changes to the VitaloJAK may compromise
product safety and/or data and as such Vitalograph cannot be
held responsible and the device will no longer be supported.
Summary of Contents for VitaloJAK 7100
Page 36: ......