ENG
17
Page
of 36
09260 Issue 1
Vitalograph VitaloJAK - Instructions for Use
for disposal of batteries.
Return shipment instructions are included in the study site manual.
Note:
Do not dispose of these products as unsorted municipal or
general waste.
6. Power Management
The VitaloJAK is powered by a battery pack, comprising 3 x AA
batteries, which should be replaced prior to configuring the device
for each recording session. Ensure the device is not attached to the
subject when changing the battery pack.
• When powered on and not recording, the Power LED is GREEN.
•
When powered on and recording, the Power LED flashes GREEN.
•
When an error state exists, the Power LED flashes RED.
If the battery pack is low on power, the device reports a Low Battery
fault, the Power LED flashes RED and the device will not operate.
See Section 7, Fault Finding Guide.
Before turning off the device, first allow any recordings to complete
to avoid data loss. Then turn off the device by turning the key so
that the red dot on the keyswitch lines up with the ‘O’ mark on the
case.
To turn on the device, turn the key so that the red dot on the
keyswitch lines up with the ‘I’ mark on the case.
7. Fault Finding Guide
Site manuals contain details for study site actions on encountering
a fault with the VitaloJAK. The following are generic fault symptoms
and possible solutions.
Problem Fault
Symptoms:
• Device does not power on
• Device powers on but does not stay on or
is not operational, e.g. can’t select a menu
item or the LCD display is corrupted.
Possible
Solutions: (In
probable order)
• The battery pack may be low/dead,
replace battery pack. A new battery pack is
required for each recording session.
• Electronics failure - contact support.
Summary of Contents for VitaloJAK 7100
Page 36: ......