8. All values displayed are expressed as BTPS values.
9. Time zero is determined using the back-extrapolated method,
from the steepest part of the curve.
10. Symptoms must take precedence over device measurements.
11. Do not expose the device to liquids other than cleaning liquids
specified.
12. Keep device dry. If the device gets wet, do not use it, and
contact Vitalograph using the contact information at the start
of this manual. Do not connect any part of this device to mains
power as there is a risk of injury especially if the device is wet.
13. The device is not intended to be used in the presence of
flammable liquids or gases, dust, sand or any other chemical
substances.
14. Service and repairs should be carried out only by the
manufacturer or by Service Agents specifically approved by
Vitalograph.
15. RF communications equipment (including peripherals such
as antenna cables and external antennas), which emit
electromagnetic fields, should be used no closer than 30 cm
(12 inches) to any part of the device, including cables specified
by Vitalograph. Otherwise, degradation of the performance of
this equipment could result.
16. The device is a Type BF applied part. The subject comes into
contact with the device, mouthpiece, SafeTway or Eco BVF
during use.
17. Take care during battery replacement. An AAA battery is a
potential choking hazard for a small child. Adult supervision
is required at all times, when a child is using the device. The
battery door, when removed, has pointed corners which may
present a risk of injury.
18. The batteries should be removed if the device is intended to be
stored or left unused for an extended period of time.
19. Only approved accessories from the manufacturer should be
used with the device. It is may be unsafe to use accessories,
detachable parts and materials not described in this document.
20. Non-Medical Electrical equipment used with the device, should
comply with its relevant IEC or ISO standard.
Summary of Contents for copd-6 4000
Page 24: ......