4. Power Management
The COPD-6 operates with 2 AAA 1.5V disposable batteries. If the
battery symbol flashes the batteries should be replaced. Replace
batteries by removing the battery door on the underside of the
device.
Note:
Dispose of used batteries safely.
5. Cleaning & Hygiene
The COPD-6 is not designed or supplied as a ‘sterile’ device.
Keep the device clean and dust free. If you suspect the device
is damaged or is measuring incorrectly, contact your medical
professional immediately.
The COPD-6 should continue to give reliable measurements for up
to three years in home use including telemedicine applications. It
should then be replaced with a new device.
5.1. Cleaning in Single Patient Environment
For single patient use, the plastic mouthpiece may be used.
Weekly cleaning of the mouthpiece, outside surfaces and flowhead
of the device is recommended. A cloth impregnated with 70%
isopropyl alcohol may be used. The plastic mouthpiece may be
washed in warm soapy water and then rinsed in clean water. The
device should be cleaned before and after an extended period of
storage.
5.2. Preventing Cross-Contamination of Subjects in Clinic use
For multi-patient use in a clinic environment Vitalograph
recommends the use of Eco BVF filters or, if these aren’t available
then SafeTway mouthpieces may be used based on the customer
own risk assessment and hygiene controls.
Before use by the next subject, the mouthpiece, outside surfaces
and flowhead of the device should be cleaned with a cloth
impregnated with 70% isopropyl alcohol. The device should be
cleaned before and after an extended period of storage. If you
suspect that a device intended for multi-patient use has become
contaminated, it should be replaced.