Instruction manual
VitalFlex Elite LTC Bed
Last update:03.16.2021
Page 15 of 34
The length of the strap of the triangle handle can be adjusted by the buckle. Select an
adjustment that allows the user to easily reach the handle when lying down (usually
between 55-70 cm measured from the upper edge of the mattress).
Make sure that the belt is securely fastened again.
The triangle handle has a durability of at least 5 years with normal use. (see embossing of
production date). It is then recommended to replace the triangle handle.
Production month
Production year
6.2
Commissioning
Connecting the nursing bed to the mains socket
Insert the mains plug into the socket.
The mains plug should always be accessible so that in an emergency the system can be disconnected
from the mains supply by pulling it out of the socket. The electric actuators are now ready for
operation.
The VitalFlex Elite LTC bed is ready for operation after it has been successfully installed
and all steps from chapter 6, paras. 6.1 and 6.2 have been observed.
Once the VitalFlex Elite has been installed, carry out a check in accordance with
Chapter 10, Para. 10.2.
Clean and disinfect the bed before using it for the first time and before each use
according to chapter 8.