2 Specifications
2.1 Digiducer 333D01 USB Digital Accelerometer Specifications
Performance
English
SI
Sensitivity
1,2
Channel A: 4.3 % FSV/g
Channel B: 8.4 % FSV/g
Channel A: 0.44 % FSV/m/s
2
Channel B: 0.86 % FSV/m/s
2
Measurement Range
3
Channel A: ±20 g pk
Channel B: ±10 g pk
Channel A: ±196 m/s
2
Channel B: ±98 m/s
2
ADC Bandwidth (-3dB)
9.3 cpm to 1374000 cpm
0.155 Hz to 22900 Hz
Frequency Range (±5%)
120 cpm to 480000 cpm
2 Hz to 8000 Hz
Frequency Range (±10%)
3
90 cpm to 660000 cpm
1.5 Hz to 11000 Hz
Frequency Range (±3dB)
3
54 cpm to 900000 cpm
0.9 Hz to 15000 Hz
Resonant Frequency
≥
1500000 cpm
≥
25000 Hz
Mounted Resonance
3
1044000 cpm
17400 Hz
Mounted Resonance
Amplification
3
200% 200%
Broadband Resolution
1
(1Hz to
10000Hz)
0.0025 g pk
0.0245 m/s
2
pk
Non-Linearity
4
≤
2%
≤
2%
Transverse Sensitivity
3
≤
5%
≤
5%
Electrical
Communication Standard
USB 2.0 Full Speed
Power Consumption
3
≤
45 mA
Internal ADC
24 bit
Supported Sampling Rates
48 kHz, 44.1 kHz, 32 kHz, 22.05 kHz, 16 kHz, 11.025 kHz, 8 kHz at 24bits /
16bits / 8bits
Case Isolated
No
Interface
USB 2.0 Full Speed / USB 1.1 Full Speed / USB Isolator
Mechanical
English
SI
Size-Hex
1.0 in
25.4 mm
Size - Height
2.6 in
66.0 mm
Weight 4.62
oz
131
gram
Mounting Thread
1/4-28 UNF
1/4-28 UNF
Mounting Torque
2 lbf.ft to 5 lbf.ft
2.7 N.m to 6.8 N.m
Sensing Element
Piezoelectric Ceramic
Piezoelectric Ceramic
Sensing Geometry
Shear
Shear
Housing Material
Stainless Steel
Stainless Steel
Sealing
Welded Hermetic
Welded Hermetic
Electrical Connector
USB Type A Male
USB Type A Male
Electrical Connection Position Top
Top
Cable (Integral) Length
9.6 ft
2.9 m
www.virtins.com 8 Copyright © 2015 Virtins Technology