Note: 

Turn the device off before replacing 

batteries.

 

 

Preparation before Taking  Measurement

 

Battery Installation

 

 
 

Battery Life and Replacement

 

 

Turn the device upside down.

 

 

Remove the battery cover by inserting the tip of your finger into the 
opening of the cover.

 

 

Install two CR2032 coin-typed batteries by following the + (positive) and

 

- (negative) polarities indicated on the battery compartment. 

Slide back the battery cover until it clicks into place.

 

Install or replace the batteries by following the + (positive) and- (negative) 
polarities indicated on the battery compartment.

 

 

When the battery level indicator appears only

 

with 1 bar left, the batteries are  nearly running 

  

                        

out, please replace new batteries timely.

 

 

 

Two new “CR2032” coin-typed batteries will last for approximately 500 
measurements.

 

 

Please use same brand and type of batteries and replace all batteries at the 
same time. Avoid using used and new batteries together.

 

 

Do not use rechargeable batteries.

 

 

Remove batteries if won't be using this device for some time, to prevent the 
device from being damaged.

 

 

Please dispose waste batteries according to applicable local regulations.

 

 
 

 

10

 

P

re
p

a

ra
tio

n

 b

e

fo

re
 T

a

k

in

g

 M

e

a

s

u

re
m

e

n

t

 

P

re
p

a

ra
tio

n

 b

e

fo

re
 T

a

k

in

g

 M

e

a

s

u

re
m

e

n

t

 

Summary of Contents for 870E

Page 1: ...870E Instruction for Use Version V1 2EN Thank you for purchasing Vion 880E Please read the instructions before use ...

Page 2: ...g Measurement Battery Installation 9 Battery Life and Replacement 10 Messages 31 Trouble Shooting 32 Technical Specifications 33 Cautions before Taking Measurement 11 Taking Measurement Operation Procedure 12 Common Wrong Operations 14 Results Display Measurement Results 15 Reviewing Results 16 BP ECG Studio System Requirements 17 Installation 18 System Authentication 20 Downloading ECG Data to Co...

Page 3: ...s of the heart disease please consult with medical professionals when feeling discomfort with your heart 6 Please follow the national requirement to dispose unit accessories waste products residues batteries etc 1 Following signs and symbols are used in the Instruction for Use on the packaging and label of the device and accessories Type CF applied part The product is with Bluetooth function ATTEN...

Page 4: ...ime these data can be transmitted to physician as reference immediately during office visits or for telemedicine purpose This device is capable of measuring and displaying up to 9 individual channels of electrocardiogram ECG rhythms Vion ECG Pen 880E 870E can also detect the presence of atrial fibrillation bradycardia tachycardia and normal sinus rhythm when prescribed or used under the care of a ...

Page 5: ...rement Do not use this device for children weighing less than 10 kgs The electrodes of this device should not come in contact with other electrically conductive parts including earth ground Do not store expose the device in the following environments Locations exposed to direct sunshine high temperature or humidity heavy dirt near water or fire and under high electromagnetic influence Do not use o...

Page 6: ...evice out of the reach of infants toddlers or children This device contains small parts such as batteries that may cause a choking hazard if swallowed by infants toddlers or children 1Static electricity may cause damage to the electronic components To prevent electrostatic damage to the device ensure your body is free of static electricity before using the device Contact with earthed objects or de...

Page 7: ...easurement records in chronicle order Display ECG waveforms to doctors for reference Your package should include the following items Main Unit Vion 880E 870E Accessories USB extension set 870E not applicable 1 Instruction for Use 2 x coin typed batteries trial use BP ECG Studio CD 870E not applicable Bluetooth 4 0 With Bluetooth 4 0 technology this device ideally provides real time display of your...

Page 8: ...he real time heartbeat During reviewing Displays the average heart rate of each measurement 3 Battery Level Indicator Shows current battery level 4 USB Connection Displays the status of USB connection 870E not applicable 5 Digits Display During ECG Recording Displays the 30 seconds count down During Reviewing Displays the sequence numbers of stored records 6 Bluetooth Connection Displays the statu...

Page 9: ...ve and negative polarities indicated on the battery compartment When the battery level indicator appears only with 1 bar left the batteries are nearly running out please replace new batteries timely Two new CR2032 coin typed batteries will last for approximately 500 measurements Please use same brand and type of batteries and replace all batteries at the same time Avoid using used and new batterie...

Page 10: ... hands are dry moisten them with damp cloth before taking measurement Do not let your right hand in contact with your body during the measurement and avoid heavily pressing the top and bottom electrodes to influence the accuracy of measuring Do not use the device in upside down position Top Electrodes Do not move or talk while taking measurement and keep the balance of your body to avoid influenci...

Page 11: ...test whether the full display is completely displayed hold the On Off button when switching on the device Taking Measurement Measurement Method A right index finger chest approximately corresponds to lead 2 1 Press the power button for 3 seconds to turn on your ECG Pen 2 Lay your right index finger on the two top electrodes of the device and hold the ECG Pen face up in your hand 3 You can determin...

Page 12: ...om electrode of the ECG Pen on bare skin about 5 centimeters 2 inches below your left nipple supported by the ribcage and press it GENTLY until you hear the click sound Draw an imaginary line from the front of your armpit downwards At the same time draw an imaginary line going up 10cm from the lowest rib on the left side of your body Place the lower electrode of the device at the point where these...

Page 13: ...electrode until you hear or feel a click Press it GENTLY until you hear the click sound Note Do not press the electrodes hardly Excessive muscle tension will cause unnecessary noise Note Please note that there must be no skin contact between your right and left hand measurement method C or hand and chest measurement methods A B Otherwise the measurement cannot be taken correctly Do Taking Measurem...

Page 14: ...easuring over or through clothing WRONG WRONG The measurement will be completed after 30 seconds and you may see the results shown on the LCD Holding the device in the wrong direction Holding the device with left hand WRONG WRONG 13 14 Note If the bottom electrode is released within 3 seconds at the beginning of your measurement the measurement will beaborted Note The latest measurement will repla...

Page 15: ...records you may Normal Result Slow Heart Rate 60 bpm Bradycardia Fast Heart Rate 100bpm Tachycardia Paused Heartbeat Irregular Rhythm Abnormal Waveform Press the button for 3 seconds to turn on the device The LCD displays the total number of records stored Press the button again and the LCD displays the average heart rate of the latest measurement and the Digits Display shows the sequence number o...

Page 16: ...Studio Insert the CD to your computer Browse the CD and double click Setup exe to start the installation Select language and continue the process 17 Read and accept the license agreement 18 BP ECG Studio BP ECG Studio Operating System Windows 7 or higher Recommended Hardware Intel Core i3 3220 3 30 GHz or higher 1 GB of RAM or more 1 GB of hard disk free space or more Screen Resolution 1280 x 1024...

Page 17: ...dio installation process Double click the icon to run BP ECG Studio The following message will be displayed BP ECG Studio Click Finish to complete the installation process Then the icon will be shown on the desktop Connect the USB extension set with your device and plug it into the USB port of your computer The system will automatically check the presence of your device BP ECG Studio 19 20 BP ECG ...

Page 18: ... to your computer Run BP ECG Studio Create a new user account by clicking 1 BP ECG Studio Fill in user information 2 Plug the connector of USB extension set into USB port of your computer 1 You need to select a user account to log in every time You may also delete the user account by clicking 2 The mandatory fields marked with CANNOT be changed once been set 21 22 BP ECG Studio BP ECG Studio ...

Page 19: ... the right column Electrocardiogram Trace Grid Color Options Click to change the grid color into green or red Scale Adjustment Zoom in Zoom out Reset the scale of signals Result Display 1 If you want to delete the data after downloading please select Delete data after download complete Note Please be aware that MD Biomedical will not be responsible for the loss of data and or information in your c...

Page 20: ...erred languages USB Connection Status Disconnected Connected Functions Menu Open the on line Instruction for Use of BP ECG Studio Delete Delete the selected ECG data from computer Provide version and copyright information User information Edit the user information Download Download data from the device to computer Configure Configure your device Print Print out the ECG BP data in a report format 2...

Page 21: ...it to sync with the computer Key Beep You can enable disable the beep sound when the device is operating Heart Beat Beep You can enable disable the beep sound during your ECG measurement Default By clicking Default you can reset your device to factory default settings Device Information You can acquire the information of ID Number and Firmware Version 27 28 BP ECG Studio BP ECG Studio ...

Page 22: ...irty please use a soft cloth or a cotton swab moistened with 70 75 alcohol to clean the electrodes Do not use alcohol to clean any other parts and use it only to clean electrodes Do not wash or immerse the device or other components in water Do not use gasoline thinners or similar solvents to clean the device or other components Note For best personal hygiene please clean the device by following a...

Page 23: ...ce with new batteries 2 Please ensure the polarities of batteries are matched with the ones indicated on the battery compartment The measurement does not start after the electrodes touching skin Press too lightly Make sure to press the bottom electrodes with sufficient force Measurement records disappear from the device storage Old records are overwritten by new ones due to storage overload Downlo...

Page 24: ...Vion ECG Pen 880E 870E ...

Page 25: ...n ECG Pen Model Vion 880E 870E Induction System single channel bipolar Bandwidth 0 05 40Hz Sampling Rate 256Hz Heart Rate Range 5 199 beats min Memory 36 memory spaces Power Source two CR2032 coin typed batteries Battery life 500 measurements with 2 new coin typed batteries Operation Temperature Humidity atmosp here 5 40 C 15 90 RH 700hPa 1060hPa Storage Transport Temperature Humidity 25 70 90 RH ...

Page 26: ... and mobile device applications during the life cycle of the device However please be aware that MD biomedical will not be responsible for the loss of data and or information in the app or personal computer 8 If any cybersecurity vulnerability or incident is detected please follow the suggestion of Messages and Trouble Shooting parts in the instruction manual first If problem still exists please c...

Page 27: ...er operation During measurement the use of the device adjacent to or stacked with other device should be avoided because it could result in improper operation In case such use is necessary the device and other device should be observed to verify that they are operating normally During measurement Portable RF communications device including peripherals such as antenna cables and external antennas s...

Page 28: ...frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to t...

Page 29: ...MD Biomedical Inc 8F No 222 Sec 4 Chengde Rd Taipei City 111 Taiwan www mdbiomedical com 35 ...

Page 30: ......

Reviews: