Replace the data cable.
Malfunction of the
chamber outlet probe
1. Flow or room temp.
exceeds allowable range
2. Data cable failure
3. Chamber run out of water
Results in Humidity less
than 80% (Invasive) or 70%
(Non-invasive) for 2 mintues
continuously after start up
for 20 minutes.
Humidification
chamber has no water
Temperature-humidity
data cable is unplugged,
or Malfunction of
temperature-humidity
data cable probe
1.Chamber is not installed
or incorrectly installed
2.Heater plate is damage
or open circuit
3. Thermal cut-out at the
bottom of heater plate is
open circuit (exceeds 118±7
℃
)
1.Heater wire adapter is
not plugged in
2.Malfunction of heater
wire circuit
Malfunction of heater
plate sensor
Contact maintenance or
local supplier
1.Securely plug in heater wire adapter
2.Replace the heater wire circuit
Add water to the humidification
chamber and restart the product
1.Securely plug in the
temperature-humidity data cable
2.Replace the
temperature-humidity data cable
1.Install chamber correctly
2.Contact maintenance
1.Check the flow or room temperature
if it’s in the range recommended.
2.Check temperature-humidity
data cable and replace if it is
damaged. Otherwise press
button to silence the alarm.
If alarm recurs, contact
maintenance or local supplier
3.Check if water-out or not.
6
4
5
8
9
10
Not
alarm
Chamber symbol
is flashing
P/N.:10017418 Rev.H Page: 17 of 22
The chamber outlet
temperature data
will display as “--”
and flash in red
Relative humidity
data on LCD will
flash in red
Summary of Contents for Inspired Medical VHB15A
Page 1: ......
Page 2: ......
Page 3: ...VHB15A Software released version 2 E 19 Oct 2017 ...
Page 4: ...3 4 Routine Inspection 5 09 05 ...
Page 5: ...15 12 11 Chapter 4 Humidifier Setup 12 12 12 ...
Page 12: ...P N 10017418 Rev H Page 07 of 22 ...
Page 13: ...P N 10017418 Rev H Page 08 of 22 110 127VAC ...
Page 16: ...P N 10017418 Rev H Page 11 of 22 Chapter 4 Humidifier Setup ...
Page 20: ...P N 10017418 Rev H Page 15 of 22 ...
Page 26: ...P N 10017418 Rev H Page 21 of 22 NOTE 1 NOTE 2 ...