P/N.:10017418 Rev.H Page: 13 of 22
5.3.2 Set Temperature
(1) Patient side Temperature setting
When the temperature front end shows “ ”, Press for 3 seconds
and the system enters the temperature setting mode. When the
temperature symbol “
℃
” is flashing on the display, the patient-side
temperature can be set. In Invasive mode, the default temperature is
39
℃
. Press or for increasing or decreasing the temperature by
0.5
℃
intervas in the range of 35 40.0
℃
; In Non-Invasive mode, the
default temperature is 34
℃
. Press or for increasing or decreasing
the temperature by 0.5
℃
intervals in the range of 30 37
℃
. After setting,
press . If in the setup process the waiting time is more than 10
seconds,the system will exit the setting mode and return to the main
page without saving all the parameters.
(2) Chamber air outlet Temperature setting
When the temperature front end shows“ ”, Press for 3 seconds
and the system enters the temperature setting mode.When the
temperature symbol“
℃
” is flashing on the display,chamber air outlet
temperature can be set. In Invasive mode,the default temperature is 36
℃
,
then the patient temperature is 39
℃
.Press or for increasing or
decreasing the temperature by 0.5
℃
intervals in the range of 34 ~ 43
℃
;
In Non-Invasive mode,the default temperature is 31
℃
, then the patient
temperature is 34
℃
. Press or for increasing or decreasing the
temperature by 0.5
℃
intervals in the range of 30~32
℃
.After setting,
press . If in the setup process the waiting time is more than 10
seconds, the system will exit the setting mode and return to the main
page without saving all the parameters.
Summary of Contents for Inspired Medical VHB15A
Page 1: ......
Page 2: ......
Page 3: ...VHB15A Software released version 2 E 19 Oct 2017 ...
Page 4: ...3 4 Routine Inspection 5 09 05 ...
Page 5: ...15 12 11 Chapter 4 Humidifier Setup 12 12 12 ...
Page 12: ...P N 10017418 Rev H Page 07 of 22 ...
Page 13: ...P N 10017418 Rev H Page 08 of 22 110 127VAC ...
Page 16: ...P N 10017418 Rev H Page 11 of 22 Chapter 4 Humidifier Setup ...
Page 20: ...P N 10017418 Rev H Page 15 of 22 ...
Page 26: ...P N 10017418 Rev H Page 21 of 22 NOTE 1 NOTE 2 ...