TS21I_IR_CO2 User Manual
Vighnaharta Technologies Pvt Ltd Page
11
of
13
3.
Press
ENTER
to select the ‘
rSt
’ function.
Unit gets reset and goes into warm-up period for 20
seconds.
Calibration process
TS21I_IR_CO2:-
TS21I_IR_CO2 needs to be calibrated once in year from factory. As year passes there is
some drift in sensor output that’s why this calibration requires.
Function names in the menu
Process Steps
1.
o1ZC
(Output 1 Zero calibration)
This is a calibration function used
to calibrate current output for
4mA.
o1SC
(Output 1 Span calibration) This
is a calibration function used to
calibrate current output for
20mA.
1.
Go to menu
o1ZC
.
2.
Measure current between
C+
and
C-
connections on
CON2
.
3.
Press
Up
or
Down
key to increase or decrease the
count to get
4mA
output across C+ and C-.
4.
Press
ENTER
key to save the count.
5.
Go to menu
o1SC
.
6.
Measure current between
C+
and
C-
connections on
CON2
.
7.
Press
Up
or
Down
key to increase or decrease the
count to get
20mA
output across C+ and C-.
8.
Press
ENTER
key to save the count.
9.
Press
ESCAPE
key to exit from Configuration mode
and remain in Idle or Scan mode.
2.
o2ZC
(Output 2 Zero calibration)
This is a calibration function used
to calibrate voltage output for
0V.
o2SC
(Output 2 Span calibration) This
is a calibration function used to
calibrate voltage output for 5 /
10V depending on voltage range
set in menu
A oP
.
1.
Go to menu
A oP
.
2.
Press
Up
or
Down
key to select 0 – 5v or 0 -10v
range.
3.
Go to menu
o2ZC
.
4.
Measure voltage between A-O/P and GND
connections on CON2.
5.
Press
Up
or
Down
key to increase or decrease the
count to get 0V output across A-O/P and GND.
6.
Press
ENTER
key to save the count.
7.
Go to menu
o2SC
.
8.
Measure voltage between A-O/P and GND
connections on CON2.
9.
Press
Up
or
Down
key to increase or decrease the
count to get 5v or 10v output across A-O/P and
GND.
10.
Press
ENTER
key to save the count.
11.
Press
ESCAPE
key to exit from Configuration mode
and remain in Idle or Scan mode.
3.
bAUd
User can set baud rate either 2400,9600,19200.
1.
Press
ENTER
to go in main menu.
2.
Use
Down
key to select ‘
bAUd
’ function.
3.
To go into ‘
bAUd
’ function, again press
ENTER
key.
4.
Use
Up
or
Down
keys to select any one baud rate