13
Open Voltage circuit
:
Less than 200mV
;
Overload Protection
:
250V DC or AC Peak Value
;
Note
:
When measuring at Range 200Ω, please short-circuit the test leads at
first to test the wire resistance, and then deduct it from the actual
resistance.
2-3-6.Diode
2-3-6.Diode
2-3-6.Diode
2-3-6.Diode
and
and
and
and
Continuity
Continuity
Continuity
Continuity
Test
Test
Test
Test
1. Insert the black test lead to “COM” terminal and the red test lead to “
”
mA
mA
mA
mA
V
V
V
V
terminal (the polarity of red lead is “+”)
2. Turn the knob to
range ,press “SELECT” button to select the
Diode measurement mode.
3. Forward Measurement: connect test leads with the diode under tested,
the red test lead connect to diode positive polarity, the black test lead to
diode cathode polarity, the reading is the approx. value of diode forward
volt drop
.
4. Backward Measurement: the red test connect to diode positive polarity,
the black test lead to diode cathode polarity ,the LCD displays “OL”.
5. The complete diode testing includes forward and backward
measurement, if the result isn’t meet the above, it means the diode is bad.
Accuracy
Range
VICTOR81
A
VICTOR81
B
Resolutio
n
200Ω
±(0.8%+5d)
0.1Ω
2kΩ
±(0.8%+4d)
1Ω
20kΩ
10Ω
200kΩ
100Ω
2MΩ
1kΩ
20MΩ
±(1.2%+10d)
10kΩ