tested voltage value has exceeded the present range limit, please select
the higher range to complete the measurement.
2. Do not measure the voltage higher than 600V. Otherwise, there is the
danger to damage the instrument.
3. When measuring the high voltage, caution to avoid the body touches the
high voltage.
Input Impedance
:
>10MΩ at Range 200mV, 40MΩ at other ranges
。
Overload Protection
:
600V DC or 600V AC Peak Value
;
Frequency Response
:(
50
~
200
)
Hz
Display
:
Average value response (RMS of sine wave)
.
2-3-3.
2-3-3.
2-3-3.
2-3-3.
DCA
DCA
DCA
DCA
1.Insert the black test lead to the COM input terminal, and red test lead to
mA
mA
mA
mA
V
V
V
V
input terminal (max 200 mV) or 10A input terminal (max 10A).
2.Turn the knob switch to Current Range, press “SELECT” button to switch to
DCA measurement mode. Connect the test leads to the tested point in
series, the current value and polarity of point connected by red lead will
be displayed on LCD simultaneously.
3. Press “RANGE” button to select auto or manual range mode.
Accuracy
Range
VICTOR81A
VICTOR81B
Resolution
200mV
±(1.5%+20d)
100uV
2V
±(0.8%+6d)
1mV
20V
10mV
200V
100mV
600V
±(1.0%+6d)
1V