6000counts Pocket Clamp Multimeter
3. Then connect the test leads across the capacitors under test.
Note:
1) When measuring capacitance with 10nF file, there may be residual reading on the screen display value. This number is the d istributed
capacitance of the test pen. It is an accurate reading and can be subtracted after measurement.
2) When the large capacitance file measures severe leakage or breakdown capacitance, some values will be displayed and unstable; when
measuring large capacitance, the reading takes several seconds to stabilize, which is normal when measuring large capacitance;
3) Please fully discharge the capacitor before testing the capacitor capacity to prevent damage to the fuse and meter.
4) Unit: 1F=1000mF, 1mF=1000uF, 1uF=1000nF, 1nF=1000pF
3-7 Frequency measurement
Input sensitivity: 1V rms; overload protection: 550V DC or AC peak (no more than 10 seconds)
Frequency measurement
1. Insert the test leads or shielded cable into the
“
COM
”
and
“
V/
Ω
/Hz
”
jacks;
2. Turn the range switch to the
“
Hz
”
position and connect the test leads or cables across the signal source or the load under test.
Note:
1) When the input exceeds 10Vrms, it can be read, but the error may be larger;
2) In a noisy environment, it is best to use a shielded cable when measuring small signals;
3) When measuring high voltage circuits, special care should be taken to avoid electric shock;
Accuracy
Range
6000counts
Resolution
10Hz
±(
0.1%+3
)
0.01Hz
100Hz
0.1Hz
1kHz
1Hz
10kHZ
10Hz
100kHz
100Hz
1MHz
1kHz
10MHz
10kHz