Summary of Contents for V1110B-THM-TEMP

Page 1: ...rform precisely as described in the documentation This system has not been designed to be used in life critical situations and must not be used for this purpose Vicon Industries Inc Tel 631 952 2288 Fax 631 951 2288 Toll Free 800 645 9116 24 Hour Technical Support 800 34 VICON 800 348 4266 UK 44 0 1489 566300 www vicon security com Document Number 8009 8300 35 00 Product specifications subject to ...

Page 2: ......

Page 3: ...njuries Alerts you to moderate dangers which if not avoided may cause minor or moderate injuries Alerts you to risks Neglecting these risks may cause device damage data loss device performance deterioration or unpredictable results Provides a tip that may help you resolve problems or save time Provides additional information Disclaimer This manual may contain misprints or product function and oper...

Page 4: ...tents Precautions i 1 Overview 1 1 1 Introduction 1 1 2 Unpacking 1 1 3 Environmental Conditions 1 1 4 Safety Information 2 2 Device 3 3 Operation 7 3 1 Procedure 7 3 2 Set Temperature 8 3 3 Calibration 8 3 4 Maintenance 8 4 Troubleshooting 9 ...

Page 5: ...Blackbody User Manual Precautions iii ...

Page 6: ......

Page 7: ... damage that may have occurred during shipment If there is shipping damage notify the carrier immediately Put the Blackbody on a steady workbench and DO NOT touch the target surface Always carry the Blackbody in an upright position 1 3 Environmental Conditions DO NOT operate the Blackbody in excessively wet oily dusty or dirty environments DO NOT operate near flammable materials Allow at least 5 9...

Page 8: ...rechamber Calibration equipment should only be used by trained personnel The power cord of the Blackbody is equipped with a three pronged grounding plug for your protection against electrical shock hazards It must be plugged directly into a properly grounded three prong receptacle DO NOT touch the IR target surface or areas surrounding the target of the instrument to avoid harm to the target surfa...

Page 9: ...3 Blackbody User Manual Device 2 Device Figure 2 1 shows the device Figure 2 1 Appearance Mounting screw hole IR target surface ...

Page 10: ...Blackbody User Manual Device Figure 2 2 Installation dimensions in mm ...

Page 11: ...5 Blackbody User Manual Device ...

Page 12: ...nt temperature 73 4º F 9º F 23º C 5º F Stability 0 27º F 10 min 0 15º C 10 min Emissivity 0 97 0 02 Power 24 VDC power supply included Size LxWxH 7 1 x 4 3 x 4 3 in 180 x 110 x 110 mm Net weight 3 1 lb 1 4 kg Operating ambient temp 32º F 104 º F 0º C 40º C 20 90 RH Install interface Supports vertical mounting ceiling mounting wall mounting and universal tripod Figure 2 3 Operation screen Temperatu...

Page 13: ...n the window below 1 3 is the set button press enter to set parameters 4 is the data move button 5 is the data decrease button 6 is the data increase button 7 is 10 LED indicators MAN light is on to indicate manual output status PRG light is on to indicate control operation MIO OP1 OP2 AL1 AL2 AU1 AU2 etc correspond to module input and output actions respectively COM light is on to communicate wit...

Page 14: ...e 3 3 Calibration To ensure the Blackbody s temperature accuracy we suggest the Blackbody shouldbe calibrated by a qualified measuring and testing institution every year 3 4 Maintenance The equipment should be maintained by a specified person and all the records should be archived Put the Blackbody into a case to avoid dust when it is not used Ensure appropriate storage temperature and humidity Cl...

Page 15: ...iption 1 No display Fuse break Replace fuse 2 Fan does not work Fan damage Return to factory Noise Add lubricant oil 3 After setting SV value PV value is not responding Heating wire burnout Return to factory SV value is close to room temperature Increase SV value Temperature controller damage Return tofactory ...

Page 16: ......