Blackbody
User Manual
Device
Table 2-1 Specifications
Parameters
Description
Temperature range
86º F
–
122 º F (30º C - 50º C)
Target diameter
2.8 x 2.8 in. (72 × 72 mm)
Temperature resolution
0.018 º F (0.01º C)
Accuracy
±0.36º F (±0.2º C (@ factory calibration,
ambient temperature 73.4º F ±9º F
(23º C
±5º F)
Stability
±0.27º F /10 min (±0.15º C/
10 min)
Emissivity
0.97±0.02
Power
24 VDC; power supply included
Size (LxWxH)
7.1 x 4.3 x 4.3 in. (180 x 110 x 110 mm)
Net weight
3.1 lb (1.4 kg)
Operating ambient temp
32º F
–
104 º F (0º C - 40º C); 20~90% RH
Install interface
Supports vertical mounting, ceiling mounting,
wall mounting and universal tripod.
Figure 2-3 Operation screen
Temperature controller
Power switch
Power socket
Cooling Fan