11
11. Technical Specifications
Type:
Forehead Thermometer V977-LA
Measuring Range:
Forehead measuring range:
34 °C (93.2°F) to 42.2°C (108.0°F).
Resolution:
0.1°C (0.1°F)
Accuracy:
Laboratory: ±0.2°C (0.4°F), 36 ~ 39°C (96.8 ~ 102.2°F)
(Ambient Temperature: 16°C (60.8°F) to 40°C (104°F))
Display:
Liquid Crystal Display, 4 digits plus special icons
Acoustic:
a. The unit is turned ON and ready for the measurement:
1 short «beep» sound.
b. Complete the measurement: 1 long beep sound.
c. System error or malfunction: 3 short «beep» sounds.
d. Temperature higher than 37.4°C (yellow or red screen
display): 10 beeps
Memory:
8 readings recall in the Memory Mode.
Backlight
a. The display will be lighted when the unit is turned ON and
will stay lit until the power button is pressed to take a
measurement.
b. The display will be lighted again for 15 seconds when the
measurement has been completed.
Operating temperature:
16°C (60.8°F) to 40°C (104°F)
Storage/transport
temperature:
-20°C (-4°F) to 50°C (122°F)
Storage/transport
humidity:
15-95% Non-Condensing
Automatic Switch-off:
Approx. 30 seconds after last measurement has been taken
Battery:
(2) AAA Batteries - at least 1000 measurements
Weight:
3.51oz. (with battery), 2.72oz. (w/o battery)
Standards:
Complies with ASTM E-1965-98 requirements
Max Power: 0.06W
Max Operation: < 20 mA
Battery: 3V