NM PLUS HYDROGEN GENERATOR
USER MANUAL
www.vicidbs.com
Version 1.06.0000
5.2. THE MOST COMMON OPTIONS
5.2.1. Remote Connections (optional)
The hydrogen generators are fitted with an optional remote control feature (USB-RS232), which allows
the user to check the status of the generator from a remote position, and to start/stop the production of
hydrogen.
The contacts used in the remote control are potentially free relay contacts. The contacts can be configured
via software as normally-open or normally-closed (see the Configuration section). The maximum voltage
and current ratings for the contacts are 1A / 48V. The pin configuration of the remote connector is shown
in the table below.
Remote connector pin configuration
Pin
Description
1 + 2
Start (12-30 VDC polarity not important)
3 + 4
Standby
5 + 6
Reaching external pressure
8 + 15
Refill water with a 24Vdc – 1A max valve. Needs gravity. Select “Yes” in the
refill menu
9 + 10
Low water level
11 + 12
Bad water
13 + 14
Change water
For part-numbers for the remote connector option please consult the list on page 11.