the pan ,hit the this key to clear the display
:
Use this key to subtract that the current weight readings is net weight.
VI.
Operation
(I) Backlight
Press key to choose the display backlight mode:
1.
AUTO: “Auto Backlight” mode. When the weight is over 5
divisions, the backlight will be switched on.
2.
ON
:
Display backlight is on all the time.
3.
OFF: Display backlight is off.
(II)Calibration
Press key, put on the weight with value same as the
screen display, when the scale is stable and the buzzer
sounds ,the calibration is finished.
NOTE: When the screen display the value of the weight,
press key for 1 second and then press ,it will
show “0- - - -- - “, then input 0-9 by , press the
key to input point, press key to figure, when the input
value flickers, put on the weight to calibration.
Press key and then press to clear the value.
Pres key for one second to exit the calibration mode.
(III)Recover the factory Set
4
Press key to return the display to zero, and then
press key to show “Centr” and before it disappear press
key to appear “RES”, press key to recover the factory
set.
(IV) Pre-set Warning
The weight range can be preset, if exceed the weight range;
the buzzer will sound every time.
(1) Pre-set low limit value and high limit value
Press the e key to shows “- - LO - -”, press the
key toto display “0----“,press key to input 0-9 .
Press key and then key to clear the value.
After setting the low limit value, the display shows”—HI—“ to
set the high limit value, the method is the same to the setting
low limit value.
After enter the low and high value, the display shows ”IN OU”,
press key to select warning out of low or high limit
value. press to select warning between low and high
value.
(2) Pre-set high limit value or low limit value
After setting the low limit value, when the display show “- - Hi -
-”, press key to show “LO Hi”, press to
select high limit value; press to select low limit value.
5