Viatran DL Series Analog Input Process Meter Instruction Manual
37
Programming the Meter (
prog
)
It is
very important
to read the following information, before proceeding
to program the meter:
•
The meter is factory calibrated prior to shipment to read in
milliamps and volts depending on the input selection. The
calibration equipment is certified to NIST standards.
•
Use
the
Scale
menu to scale the process input (e.g. 4-20 mA). A
calibrated signal source is not needed to scale the meter.
•
Use
the
Calibrate
menu to apply a signal from a calibrator or a
flowmeter.
•
The meter is a single input meter with dual-scale capability.
The
Program
menu contains the
Scale
and the
Calibrate
menus.
Note: The
Scale
and
Calibrate
functions are exclusive of each other.
The meter uses the last function programmed. Only one of these
methods can be employed at a time. The Scale and Calibrate functions
can use up to 32 points (default is 2). The number of points should be
set in the Advanced menu under the Multi-Point Linearization
(
Linear)
menu selection prior to scaling and calibration of the meter, see page
74 for details.
If the dual-scale level feature is selected in the
Setup
menu, the
Scale 1
and
Scale 2
menus are enabled for PV1 & PV2 respectively.
The process input may be calibrated or scaled to any display value
within the range of the meter.
Program Menu for Single Scale Process
Program Menu for Dual-Scale Level Applications
Additional parameters, not needed for most applications, are
programmed in the
Advanced
Features
menu; see
Advanced Features
Menu
, page 67.