19
ELITE
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The ELITE is intended for use in the electromagnetic environment specified below. The customer or the user of
the ELITE should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
level
Compliance
Electromagnetic
Environment-
Guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
+6 kV contact
+8 kV air
+6 kV contact
+8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered
with synthetic material,
the relative humidity
should be less than 30%.
Electrical fast transient/
burst
IEC61000-4-4
+2 kV for power
supply lines
+1 kV for input/
output lines
+2 kV for power supply
lines
+1 kV for input/output
lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
+1 kV differential
mode
+2 kV common
mode
+1 kV differential mode
+2 kV common mode
Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
lines
IEC 61000-4-11
< 5%
U
T
(>95% dip in
U
T)
for 0.5 cycle
40%
U
T
(60% dip in
U
T) for
5 cycles
70%
U
T
(30% dip in
U
T) for
25 cycles
<5%
U
T
(>95% dip in
U
T)
for 5 sec
< 5%
U
T
(>95% dip in
U
T) for
0.5 cycle
40%
U
T
(60% dip in
U
T) for 5
cycles
70%
U
T
(30% dip in
U
T) for 25
cycles
<5%
U
T
(>95% dip in
U
T) for 5
sec
Mains power quality
should be that of a
typical commercial or
hospital environment. If
the user of the ELITE
requires continued
operation during power
mains interruptions, it is
recommended that the
ELITE be powered from
an Uninterruptable
power supply or its
battery.
Power Frequency (50/60
Hz)
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should
be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Note:
U
T is the a.c. mains voltage prior to application of the test level.