![VIASONIX FALCON ABI Plus Service Manual Download Page 30](http://html.mh-extra.com/html/viasonix/falcon-abi-plus/falcon-abi-plus_service-manual_1011580030.webp)
Falcon Service Manual
___________________________________
30
c) Unscrew the external nut holding the internal temperature cable as shown below and remove
the internal temperature cable.
5.9
Remove a Check Valve
Each pneumatic channel has one mechanical check valve. The role of the check valve is to ensure
that the cuff pressure does not exceed the approved limits. The check valves are directional, and
should be connected to the tubing according to the arrow direction which appears on the check
valve as shown below. To replace a check valve simply disconnect the valve from the silicon tube
by pulling it away as shown below. When replacing the check valve ensure that the arrow on the
valve is as shown below.
VS0152, rev 1.3
Page