Falcon Service Manual
___________________________________
25
5.7
Release the CW Doppler box
a) The CW Doppler box is located along the internal side of the Falcon front panel across the 3
Doppler connectors (not applicable for Falcon/ABI+). A flat cable coming out of the box
connects it to the Falcon main board. The flat cable is connected with a wire tie to internal side of
the right panel to avoid potential vibrations. Cut the wire tie on the flat cable as shown below.
b) Pull on the flat wire to disconnect it from the Falcon board as shown below.
VS0152, rev 1.3