Analytical balances 

series XA/X 

 
 
 

 

 

 

 

 

 

 

Vetek AB  

Box 79, Industrivagen 3    

Tel +46-176 208 920  

www.vetek.com  

760 40 Vaddo, Sweden    

Fax +46-176 208 929  

e-mail [email protected] 

Summary of Contents for X Series

Page 1: ...Instruction number LMI 31 03 05 09 ENG Analytical balances series XA X Vetek AB Box 79 Industrivagen 3 Tel 46 176 208 920 www vetek com 760 40 Vaddo Sweden Fax 46 176 208 929 e mail info vetek com ...

Page 2: ... 2 MAY 2009 ...

Page 3: ...ION 22 8 1 Automatic balance calibration 23 8 2 Calibration test 24 8 3 Manual calibration 25 8 4 Calibration report printout 27 9 SETTING PRINTOUTS FOR GLP PROCEDURES 28 10 SETTING TIME AND DATE 28 11 SETTING THE PARAMETERS 31 11 1 Setting filter 31 11 2 Value release 32 11 3 Set the display refreshing time 32 11 4 Set autozero working 32 11 5 Last digit 32 11 6 Negative 32 12 FUNCTIONS IN USING ...

Page 4: ...PRINTOUTS 61 18 1 Standard printout 61 18 2 Non standard printout 62 19 COOPERATION WITH PRINTER OR COMPUTER 67 19 1 Connections 68 20 COOPERATION WITH LABEL PRINTERS CITIZEN CLP 521 69 21 COOPERATION WITH PRINTER EPSON 74 22 WEIGHING LOADS UNDER THE BALANCE 75 23 CONNECTING ADDITIONAL KEYS 76 24 LIST OF COMMUNICATIONS COMPUTER BALANCE 76 25 COMMANDS ABOUT ERRORS 79 ...

Page 5: ...0Hz AC 11V AC Calibration internal automatic Display graphic with backlight Max ambient humidity 80 Min ambient humidity 35 XA 100 X XA 160 X XA 220 X XA 310 X Max capacity 100 g 160 g 220 g 310 g Min capacity 10 mg 10 mg 10 mg 10 mg Readability 0 1 mg 0 1 mg 0 1 mg 0 1 mg Tare range 100 g 160 g 220 g 310 g Repeatabilty 0 1 mg 0 1 mg 0 1 mg 0 2 mg Linearity 0 2 mg 0 2 mg 0 2 mg 0 3 mg Pan size Φ 1...

Page 6: ...essary to correct performance of balance elements off the box After replace balance in place of use put scale and rest of elements on The balance is supplied with 230 V AC 11V AC supplier The balance with Φ 100 mm pan open side doors of weighing chamber inside the chamber place lower shield of weighing chamber 3 inside the chamber place anti breeze shield 2 put the balance pan inside the ring 1 cl...

Page 7: ...y pin of power adaptor should be connected to socket in the back of balance casing The balance with pipettes calibration set slide chamber side doors off 8 inside the chamber put bottom ring 8 put glass ring on bottom ring 7 replace pan inside glass ring 6 put upper ring on glass ring 5 put steam protection container on upper ring 4 put container to pipettes calibration inside the container 3 put ...

Page 8: ...he display put the container on the pan and after stabilization press ESC ZERO TARA 3 2 Set level up After replacing the balance in the place of usage the balance should be leveled The correct leveling is shown on the level indication installed at the rear of the balance The air bubble should be in the centre position of the indication 3 3 Warmin up Before measurements user should wait untill the ...

Page 9: ... counts pieces weights dosages determines deviations of the standard mass weights animals detemines liquids and solids density making mixtures up to recipes statistics of measurements series recipes Measure units and particular functions can be inaccessible for user It is possible to adapt the balance to individual needs and access functions and units which are necessary at this moment It is possi...

Page 10: ...1 load mass and quantity of pieces 2 measure unit 3 the result is stable 4 line of max range of the balance 5 work mode 6 date 7 time 8 precise ZERO 5 2 Keyboard Each key is dual function key Particular function can be done through User also can move in the balance menu ...

Page 11: ...ey sends information to external instrument PRINT or confirms parameter value or function ENTER Key ESC ZEROTARA sets indication to zero Calibration Menu of balances Navigation bar 5 3 Connections 1 power adapter socket 2 PS keyboard connector 3 RS 232 port 4 additional display socket Draw 4 Sockets of the XA X balances 6 USER MENU There are 9 groups in user menu Each group is named by P letter Na...

Page 12: ... off 08 Last cal print 0 0 off P3 Date Time 01 Date format 0 DA MO YR 02 Time format 0 24 h 03 Time function 04 Date function 05 Display time 1 yes 06 Display date 1 yes P4 Readout 01 Filter 3 normal 02 Value relase 1 fast rel 03 Disp Refresh 1 0 08s 04 Autozero 1 on 05 Last digit 1 always 06 Negative 1 disabled P5 RS 232 01 Baud rate 1 4800 02 Parity 0 none 03 Data bits 2 8 bits 04 Stop bits 1 1 ...

Page 13: ...ele Hk 1 enabled 09 Taele S 1 enabled 10 Taele T 1 enabled 11 Momms 1 enabled 12 Grains 1 enabled 13 Newtons 1 enabled 14 Ticaal 1 enabled 15 Custom 1 enabled P8 Work modes 01 Parts count 1 enabled 02 Checkweighing 1 enabled 03 Dosages 1 enabled 04 Deviations 1 enabled 05 Animals 1 enabled 06 Density 1 enabled 07 Recipes 1 enabled 08 Statistics 1 enabled P9 Others 01 Setting ID function 02 Autopri...

Page 14: ...rom the balance memory noted changes sets in the balance memory eg Date time user number texts Menu graphic version Press the SETUP key to display main menu of the balance display I Select the submenu whose contents is displayed after pressing the RIGHT ARROW key display II Draw 5 Menu 1 main menu number 2 indication of the function selection 3 function name 4 currently used function 5 submenu num...

Page 15: ...n the menu selects submenu which is activated submenu contents is presented on the display move one level up to main menu Resignation parameter changing 6 1 2 Return to weighing function Introduced changes are recorded after return to weighing mode and confirm changes Press the ESC key many times If following question appears on the display press ENTER confirm or ESC cancel Draw 6 Return to weighi...

Page 16: ...rd has its equvivalent on the PS keyboard for functions Description keyboard Switch on off the balance display Move to the balance menu Selects work mode Selects measure unit PRINT TARE for Direktion keys Move up Move to level up Sets selected parameter Move down ...

Page 17: ...ance manu and control the balance work The list of the commands is at the end of the manual 6 2 User menu The menu is presented in p 5 Press the Setup key in weighing level Main menu is presented on the display Select the submenu which is modificated Draw 7 Main menu submenu selection If the menu is modificated press the RIGHT ARROW key Selected menu appears on the display Select what will be chan...

Page 18: ...nce with p 7 1 Before measurements load the pan and check if the balance show precise zero displayed in down left corner of the display only if the parameter P4 06 Autozero has the value 1 yes and check if the measurement is stable is displayed in right up corner of the display If the balance does not show zero press the key If the conditions are unfavourable no stable result lines appear on the d...

Page 19: ...ter software PIPETTES user can calibrate pipettes pipettes calibration stand Function is described in further part of manual Independently on pipettes calibration for this function steam protection ring must be installed on see point Unpacking balance Container of steam curtain minimizes errors of measurement when liquid evaporates during weighing Before pipettes calibration pour water to steam pr...

Page 20: ...thorised to sets and the balance functions determined by administrator Setting password and access authorization After set the password and access parameters parameter P9 13 Password protection write the password for administrator If the admistrator password is different from 0 the program wants administrator password during enter for the parameter P9 13 Password protection Every enter this parame...

Page 21: ... set up on YES user can perform the calibration and calibration report Stars If it is settled on YES during start the balance up password is hidden under starts Administrator password Write the password for administrator max 6 digits and user Each admistrator has access to all functions in the balance User has access to balance possibilities in accordance with above description Please remember the...

Page 22: ...ure inside the balance changes more than 0 8 C Kind of calibration Internal automatic calibration started if temperature changes started if the time changes Internal manual calibration started by the balance keyboard Calibration made with external weight with declared mass which cannot be modificated with any mass which should be given before the calibration process for not legalized balances only...

Page 23: ... untill the measurements are performed If the T O key is pressed the calibration process is stopped Set up automatic calibration Draw 12 Set up automatic balance calibration 1 main menu number 2 function selection factor 3 function name 4 name of actual activity 5 selects factor to autocalibration time temp 6 declaring autocalibration time 7 value of factors for autocalibration 8 value of time for...

Page 24: ...alibration determine factor which decides about start automatic internal calibration 0 non non of the factors causes start of the calibration 1 time calibration in relation to time determined in p 07 2 temperature calibration in relation to changes of surroundign temperature 3 both calibration in relation to changes of time and temperature 07 Automatic calibration time Determination of time automa...

Page 25: ...rds results of the calibration in the memory and returns to weighing mode Pressing the ESC key stops the calibration process If during the calibration load is on the pan display show order about error The calibration process is stopped After take load off the calibration process is finished If the function DRH is active user cannot stop the internal calibration process 8 3 2 External calibration T...

Page 26: ...ation mass The first digit flashes and it can be changed Draw 14 User calibration declaring value of weight Record new external mass by functional keys in accordance with p 5 1 1 of the manual Confirm the mass The balance starts calibration and shows orders on the display The balance determines mass of empty pan and shows order to put this mass After put the weight on the pan confirm by the Enter ...

Page 27: ...the report is generated and sent automatically Draw 15 Submenu calibration A content of report depends on setting in submenu GLP All options with YES attribute are printed Draw 16 Submenu GLP setting Apart from information settled in menu group the report contains calibration mass remembered by balance after last calibration description Old calibration mass determined during actual calibration des...

Page 28: ...by the balance keyboard or the PS 2 keyboard For the rest select 1 no do not print during report 0 yes print during report Main view of the GLP submenu is presented on drawing 19 page 24 If you use the PS 2 keyboard see what dependences are between the balance keyboard and PS 2 keyboard p 5 1 2 10 SETTING TIME AND DATE The balance has real time clock whose parameters can be modificated Enter the m...

Page 29: ... 24 h After selection press the ENTER to confirm 12 h form is distinguished by the letters PM or AM on the printouts 03 Time Enter setting the parameter 03 Time by the RIGHT ARROW key in accordance with below scheme Draw 19 Submenu Date Time setting time Replace the marker next to the value which will be changed Hour Minute Second Confirm with the RIGHT ARROW key Change the numerical values with U...

Page 30: ...R key The balance returns to submenu P3 Date Time and displayed time changes After setting time return to weighing mode in accordance to p 5 1 1 of the manual 04 Date Set the parameter 04 Date with the RIGHT ARROW key In accordance with previous description 03 Time set actual date After setting date return to weighing mode as it is presented in p 5 1 2 of the manual Draw 21 Submenu Date Time setti...

Page 31: ...eans of parameters in group P4 Readout Draw 22 Submenu Readout internal setting 11 1 Setting filter Dependly on conditions set the filter If the conditions are conductive set the filter as very fast value of the parameter 01 Filter 1 and if the conditions are bad vibrations draught set the filter as slowly or very slowly value of the parameter 01 Filter at 4 or 5 Efficience of filter is different ...

Page 32: ...ise indications programmable function AUTOZERO is in the balance This function controls automatically and corrects zero indication of the balance If the AUTOZERO function is active each measurement starts at precise zero every time In special cases this function disturbes in the measurements eg when the load is put on the pan very slowly pouring substance In this case correcting system of zero ind...

Page 33: ...3 when stable 07 Interval Interval the balance sends the result from the display It is counted on base on form for the parameter x 0 1 s time yestu interval Value from 1 to 9999 can be written 08 Min mass Min mass for automatic work RS 232 Following measurement is sent when the result is lower than min aftre taking the load off the pan 09 Print only stable 0 no 1 yes 10 Type of printer Epson or st...

Page 34: ... switching between units For units described as 09 Taele Hk 10 Taele S 11 Taele T there are following dependences If all of them have attribute 1 yes the balance show only first of them 09 Taele Hk If the measurement is done in units 11 Taele T set the attribute 0 no for two previous Enter group of the parameters P7 Units according to p 5 2 7 Draw 24 Measure units setting After set proper values o...

Page 35: ... the parameters can be done according to p 5 1 1 of the manual 16 OTHER PARAMETERS User can set parameters have influence on work with balance in group of the parameters P9 Others eg beep signals etc Enter submenu group P9 Others the same as in p 14 01 Setting ID it includes 6 digits 6 codes which can be used during printouts for product specification operator batch etc 02 Automatic printout ID fo...

Page 36: ...b angielski 09 The balance number It is only information about factory number of the balance 10 Number of the program It is information about program of the balance 11 Printout of the parameters if the function is active the balance parameters in user menu are printed User gives numbers of the parameters which should be printed Draw 26 Submenu Others printing setting After confirmation parameters ...

Page 37: ... set up User must know all parameters and computer operation very well 13 Password protection this submenu contains options about access password for administrator and user see 6 1 17 USING WORK MODES 17 1 Counting pieces of the same mass It can be done after write singular piece mass Write singular piece mass Determine singular piece mass on base of standard quantity Element selection from date b...

Page 38: ...tions Select MO Weighing Press the RIGHT ARROW display show stage of weighing 17 1 2 Counting through determine singular element mass from the standard batch Start the procedure of counting pieces in accordance with p 16 1 1 it does not matter which mass in the field 01 Select 07 Start and press the key RIGHT ARROW In the counting pieces function press the key Units Dialog window appears on the di...

Page 39: ...s are counted in relation to singular mass This way mass of singular piece can be determined on base of batch standard The are four conditions of AKD Automatic Correction of Preciseness in the balance software 1 quantity of pieces after adding must be higher than it was previously 2 quantity of pieces after adding must be less than twice quantity which was on the display before adding 3 actual qua...

Page 40: ...the Enter record name and Enter record value Next to name singular piece mass should be introduced It can be remembered using 02 Recall sample 17 1 3 Select piece from data base Active function of counting pieces as it is shown on below scheme Draw 32 Select piece from data base Select piece from data base Start counting pieces ...

Page 41: ...graphs and controled The function activation Draw 33 Checkweighing the function activation Display Draw 34 Checkweighing display view 1 result 2 bargraphs 3 function name 4 difference between mass of weighed load and middle of tolerance field HI LO 5 value of low LO and HI high limit 6 graphs which presents weighing range ...

Page 42: ...setting parameters without changes Checkweighing with stage base Checkweighing in RANGE BASE which contains 500 records Name of the product for each record max 10 alphanumerical signs Value of upper range for each record Value of lower range for each record Programming stage base Draw 35 Checkweighing submenu view Move into Stage base and press RIGHT ARROW Select record number and press RIGHT ARRO...

Page 43: ...W Draw 37 Checkweighing stage base programming values Enter name for selected record product name Enter value of upper range Enter value of down range Confirm values with ENTER press twice Draw 38 Checkweighing start work with checkweighing ranges ...

Page 44: ...oduct recorded in range base 8 difference between result and the ranges 9 graphic view which place weighed mass is 10 used load of the balance Selecting other product from stage base Press SETUP during checkweighing Display shows dialogue windows move to range base and select other product or enter other values for up and down range if you use range base 17 3 Dosage During dosage pouring load mass...

Page 45: ... function Draw 40 Dosage activation Display Draw 41 Dosage display 1 mass which should be added 2 graphs 3 function name 4 TR reference value mass which is declared see drawing 38 M3 01 Reference mass 5 WGH mass on the pan ...

Page 46: ... in percentages Following functions dosage weighing statistics can cooperate with deviation function Activation of the function Draw 42 Percents activation Display Draw 43 Percents display 1 percentage value proportion of the mass on the pan and standard mass 2 function name 3 REF reference mass see drawing 40 M4 01 4 WGH mass on the pan ...

Page 47: ...istics and change the attribute NO into attribute YES Confirm this option and press the key Enter Draw 44 Percents cooperation with other functions 1 percentage value relation of the load on the pan to reference mass 2 stable measurement sign 3 function name 4 REF reference mass 5 WGH mass on the pan 6 graph which presents weighign range where the weighing range is 7 statistics N 0 no measurements...

Page 48: ...esult can be printed Return to statistics submenu and higher levels key ESC Draw 45 Percents cooperation with other functions Statistics Reference mass In this software reference mass can be defined by weighing standard Press F key in main menu window The software starts procedure and displays special command Please follow the commands When this procedure is finished the software returns to main w...

Page 49: ...er filter the shorter time of measurement STAGE Value in actual scale intervals is value the result must be below The result of weighing must be smaller than value of actual scale intervals in order to do following automatic measurement AUTO START Automatic start up following measurements STATISTICS Statistics counting for particular subjects START Start measurements up ...

Page 50: ...alytical balance there are Specific Gravity Measurement Kit Draw 47 Specific Gravity Measurement Kit Components of the kit 1 Beaker stand 2 Pan stand 3 Float 4 Beaker 5 Termometer clamp 6 Termometer 7 String 8 Float hook 9 Top pan 10 String 11 Bottom pan 12 Attachments ...

Page 51: ...ough RS 232 to printer or computer after pressing PRINT key 17 6 2 Density of solids Density of solids can be determined in one of 3 different liquids WATER destiled water ALCOHOL spirit 100 0 1 at 20 C OTHER other liquid with known density Measurement of density of solids is based on comparison sample mass in air weighed on top pan to the same sample mass in the liquid on bottom pan The program c...

Page 52: ...me max 10 signs 3 mass which should be measured for actual weighed component WGH 4 quantity of components which is weighed in the mixture IC 5 components mass already weighed SUM The function activation Draw 48 Recipes internal setting parameter 01 Hints after set the parameter at YES the balance displays names and singular components mass recorded in the parameter 04 Recipe on the graphic display...

Page 53: ...r 06 Statistics switch on YES or switch off NO statistic counting Attention Statistics counting refers only to total mass of prepared mixtures singular components mass are not counted parameter 07 Start enter work modes Recipes Information on the graphic display for recipes Draw 49 Recipes functions 1 mass which is actually on the pan 2 stable measurement sing 3 function name 4 settled mass of the...

Page 54: ...corded components and their mass in the balance memory Write names and components mass in the parameter 04 Recipe Remember about the dependences there cannot be more than 10 signs confirm each name by the ENTER key and write mass which will be in the mixture Draw 51 Declared recipes ...

Page 55: ...t component mass in the WGH Press the UNITS key Mass of component 1 is recorded in the balance memory The information on the display changes component 2 mass WGH IC 1 SUM Information on the display is settle do zero Repeat it for all components After weighing last component and write its mass to the balance memory the UNITS key total mass of mixture and prompts to following steps are displayed Pro...

Page 56: ...erform measurement series 6 enter the parameter 06 Statistics again 7 enter the parameter 06 02 Results 8 to print results press the key PRINT 17 8 Statistics Activation Draw 52 Statistics function activation Results of previous statistics should removed after function activation It is realized through option M8 01 Cancel All statistic date are actualized after write following measurement to the b...

Page 57: ... 2 measurement number in measurement series 3 sum of all weighed components in measurement series 4 average mass of weighed components in the series 5 mass of the lightest component in measurement series 6 mass of the heaviest component in measurement series 7 difference between the lightest and the heaviest component in measurement series 8 value of counted standard deviation 9 value of variation...

Page 58: ...ental conditions Temperature of environment tip and liquid should be 20 C 25 C stabilized during weighing in range of 0 5 C Moisture about 50 75 Use distillated water only in a calibration process Function activation Draw 54 Calibration of pipettes activation ...

Page 59: ... V1 2max Error of capacity Vmin Temperature of liquid oC Atmospheric pressure hPa Moisture Amount of samples Display view Draw 55 Calibration of pipettes display 1 result 2 function is active 3 capacity checked 4 average capacity 5 precision error 6 amount of samples measurements Final result of the procedure is error of precision ES standard deviation sr and repeatability error CV ...

Page 60: ... 60 After each check a result is presented on display Draw 56 Calibration of pipettes summary A report is printed after pressing PRINT key Draw 57 Example of a report of in calibration procedure ...

Page 61: ... printout It includes result of weighing and all variables which have attribute YES in GLP submenu In User and Project fields names should be written Draw 58 Declaration of variables to printout submenu GLP Example of standard printout Draw 59 Example of standard printout all options settled on YES printed ...

Page 62: ...rinted And then write text in the lines 1 40 It is recomended to use PC keyboard what is simplier and faster way Non standard printouts can overlap each other Printout 1 Start 1 Printout 1 Stop 40 Printout 2 Start 20 Printout 2 Stop 40 Non standard printout can be created by Edition of the printout During manual writting give all special signs as CRLF tabulator etc If function Printout Edition is ...

Page 63: ...ut of singular sign N Actual net mass in basic unit d Actual date t Actual Time i The balance number R The program number P The Project number U The user name F Actual function name work mode C Date and time of last calibration K Kind of last calibration I Deviation of last calibration 1 Code 1 2 Code 2 3 Code 3 4 Code 4 5 Code 5 6 Code 6 Variables dependent on used work mode Variable Description ...

Page 64: ...which value depends on the mode V Mass in actual unit Value connected to work mode eg counting pieces for mode Counting pieces or deviation from standard mass in for mode Deviation Special signs used to create special printouts Singular sign c CRLF r CR n LF t Tabulator s Skip to next stringu 0 End of the printout Each text Text 1 89 Text 80 can include max 8 signs letters digits special signs spa...

Page 65: ... c 23 Text 14 z a ć 1 1 2 24 Text 15 5 5 0 g Example 2 Zakład Mechaniki Precyzyjnej RADWAG Date Time Moad mass Signature actual work mode Set following texts and write 8 signs in each of them untill it is finished Parameter number Text 1 2 3 4 5 6 7 8 25 Text 16 Z a k ł a d M 26 Text 17 e c h a n i k i 27 Text 18 P r e c y z y 28 Text 19 j n e j R A 29 Text 20 D W A G c D 30 Text 21 a t a d c 31 T...

Page 66: ...g sign is cancelled Confirm the text On PS 2 keyboard Press F2 to enter main menu Press F3 to set parameters indications next to group P6 Printouts and press F2 to enter menu group and then select parameter Press F2 to activate the procedure of writing the text By means of keyboard write the text max 8 signs and confirm by Enter Repeat this procedure for the rest of the texts Description of the co...

Page 67: ...n 18 2 3 Select non standard printouts if STANDARD printout is selected there will be only result and variables declared in the GLP see p 17 1 Standard printout drawing 60 Declaration of the variables to printout submenu GLP If non standard printout is printed select kind of the printout 1 4 and give the beginning and end of the printout 19 COOPERATION WITH PRINTER OR COMPUTER To send the informat...

Page 68: ...aw 64 connections balance computer The balance connection DB 9 F The computer connection DB 9 F with control of sending date Balance Computer 2 RxD 3 TxD 3 TxD 2 RxD 4 DTR 6 DSR 5 GND 5 GND 6 DSR 6 DTR 7 RTS 8 CTS 8 CTS 7 RTS ...

Page 69: ...ct a proper number of characters for each variable charts 2 3 and 4 After giving it an original name save the label on a computer hard drive alphanumeric characters Download the label to the CITIZEN CLP 521 printe Set RS232 baud rate in the printer to 9600 bits per second example of label named Etykieta01 Draw 65 Label example Design special printout which enables labels according to the design Se...

Page 70: ...02L c Start of the label rlabelname c Label name X c Start of variables edition 02U01NN c Variable 1 NN variable symbol 02U02NN c Variable 2 NN variable symbol 02U03NN c Variable 3 NN variable symbol 02UnnNN c Variable nn NN variable symbol 02fnnn c Paper feeding nnn mm depending on the label size E c End of variables editio 02Ennnn c Print nnnn labels 02G c End of the label An example of an inter...

Page 71: ...arameters Now you can connect the balance with the printer a cable diagram in the users manual Check the setting of transmission parameters both in the balance and in the printer They should be the same Place a load on the weight pan and after stabilizing press PRINT The designed printout will be sent to the printer and three identical labels will be printed out The printout apperance Draw 68 Prin...

Page 72: ...perator number F X Name of present operation work mode C 25 Date and time of last calibration K X Type of last calibration I 16 or 18 Deviation in the last calibration 1 6 Code 1 2 6 Code 2 3 6 Code 3 4 6 Code 4 5 6 Code 5 6 6 Code 6 V 16 or 18 Mass present unit or a value connected with present operation mode eg Number of pcs for pcs counting or deviation of standard weight in for deviation mode ...

Page 73: ...g b 14 Threshold i 14 Liquid Mass density measurement p 14 Procedure c 14 Temperature a 16 Liquid mass den sity v 16 Plunger volume TABELA 4 Zmienne statystyczne występujące we wszystkich modach oprócz ważenia podstawowego Zmienna Ilość znaków Opis zmiennej n 7 Measurment number x 16 Avarage value S 16 Sum m 16 Minimal value M 16 Maximal value D 16 The difference between minima and maximal value s...

Page 74: ...ing to below tables with Polish signs ą ć ę ł ń ó ś ź ż CP 852 A5 86 A9 88 E4 A2 98 AB BE Ą Ć Ę Ł Ń Ó Ś Ź Ż CP 852 A4 8F A8 9D E3 E0 97 8D BD instead of polish signs write their code equvivalents To have the paper cut after each printout following command should be entered 1D 56 41 08 C Connection scheme balance printer Citizen balance printer Epson Draw 69 Scheme of connection balance printer Cit...

Page 75: ...ance There is a hook visible which can be used to suspend a sample Hang the load on the look in the hole under the balance the hook is a part of standard equipment Note 1 The suspension eye is directly connected to the weighing mechanism Take great care not to move or rotate the eye as this will damage the weighing mechanism 2 The mass of all suspended accessories should be set to zero by pressing...

Page 76: ...mand PC CR LF all recorded information in commands in the balance programme are sent from the balance Function SEND THE RESULT IN BASIC UNIT Command S CR LF result is sent from the balance in basic interval after stability Function SEND RESULT IN BASIC UNIT IMMEDIATELY Command SI CR LF Function SEND THE RESULT IN ACTUAL INTERVAL Command SU CR LF result in actual unit is sent from the balance after...

Page 77: ...on WRITE NEW OR CHANGE TIME IN THE BALANCE Command PD CR LF the balance sends settled time or this time is changed Function WRITE ACTUAL WORK MODE Command PM CR LF Function SEND SETUP Command PS CR LF all balance setup is sent printout of the parameters Function SOUND SIGNAL BEEP Command B CR LF sound beep is switched on Function SEND LAST ERROR CODE Command ER CR LF last order of the error is sen...

Page 78: ...nction SWITCH ECHO OFF FOR THE KEYBOARD Command E0 CR LF keys codes are switched off Function SWITCH ECHO ON FOR THE KEYBOARD Command E1 CR LF Function SWITCH THE BALANCE OFF Command O0 CR LF the same as ON OFF Function SWITCH THE BALANCE ON Command O1 CR LF the same as ON OFF Function SWITCH AUTOZERO OFF Command A0 CR LF Function SWITCH AUTOZERO ON Command A1 CR LF If command which is not listed ...

Page 79: ...mass recorded in the balance memory higher than difference 1 Piece 1 Div 2 10 Singular mass value in counting pieces function less than actual scale interval Piece 10 Div 2 11 Mass on the pan during determining mas sof singular piece in the function of counting pieces less than 10 actual scale intervals Ref 1000 Div 2 12 Value of reference mass in the function deviations is less than 1000 actual s...

Page 80: ... 80 ...